15
Switching the pump on in subsequent power cycles
Subsequent power-up sequences will jump from the start-up screen to the home screen.
l
The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a
fault is found, an error code is displayed.
l
The pump displays the start-up screen with the Watson-Marlow Pumps logo for three seconds
followed by the home screen
l
Start-up defaults are those in place when the pump was switched off last
Check that the pump is set to operate as you require it. The pump is now ready to operate.
All operating parameters may be changed by means of key-presses (see
).
Power interruption
This pump has an auto restart feature (which only affects manual mode, network mode and memodose
operation) which, when active, will restore the pump to the operating state it was in when power was
lost.
Stop/start power cycles
Do not power up/power down the pump more than 12 times in 24 hours, whether manually or by
means of the auto-restart facility (which only affects manual mode, network mode and memodose
operation). We recommend remote control where a high frequency of power cycles is required.
If the Auto restart feature is enabled it may cause the pump to start as soon
as power is turned on.
Auto-restart only affects operation of manual mode, network mode and
memodose mode.
If Auto restart is enabled the "!" Symbol is shown on the screen to warn users
that the pump may operate without any manual intervention (pump resumes
with previous settings).
Do not use auto restart for more than 12 times in 24 hours. We recommend
remote control where a high number of starts is required.
If the pump is configured to network mode or analog mode it will respond to
remote commands at any time including immediately after power on. The "!"
Symbol is shown on the screen to warn users that the pump may operate
without any manual intervention (e.g. a remote setpoint could start the pump
with no key press required).
48
m-630dun-en-08.2