List of parameters
Param
No.
EDS Values
Read/Write
Default
Value
(EDS)
Parameter
Range
Comments
1
SetFlowCal
Write
15120
1 -
2147483647
Use to set the tube flow
calibration value. The flow
calibration value is set in
μL.
2
SetSpeed
Write
100
1 - 2200
Speed is set in deci-rpm.
Max speed depends on
head type. See
3
SetSpeedLimit
Write
2200
1 - 2200
Speed is set in deci-rpm.
Max speed depends on
head type. See
4
SetFailsafeSpeed
Write
100
1 - 2200
If the failsafe is enabled the
pump will run
continuously at the
selected speed in the event
of a communications loss.
5
SetFailsafeEnable
Write
0
0-1
(False/True)
Enabled the failsafe speed.
If disabled the pump will
stop in the event of a
communications loss. If
enabled the pump will run
at the speed set in the
“SetFailsafeSpeed”
parameter
6
SetReverse
Write
0
0-1
(False/True)
If set the pump will run
anti-clockwise. Pump
defaults to clockwise
rotation
7
Run
Write
0
0-1
(False/True)
Set to 1 (true) to allow the
pump to run. 0 will stop
the pump. Note that pump
enable needs to be set
m-730en-en-08.1
55