Replacing the passive-preheater assembly in a CM-30S installed in a UPC
2
system
October 7, 2014, 715004178 Rev. B
115
5. If you are replacing the passive preheater in position 8, route its tubing
above the upper tubing guide post and then below the middle tubing
guide post adjacent to the left-hand column plate.
Requirement:
When replacing the passive preheater in position 5, 6, 7,
or 8, use the right passive preheater assembly.
6. Straighten the valve fitting end of the tubing enough to thread the
tubing through the opening at the back of the valve drip tray.
Warning:
To avoid spills, do not remove the valve's drip tray when
routing the passive-preheater tubing. All drip trays must remain
in place during the CM-30S's operation and maintenance.
Middle tubing guide post
Upper tubing guide post
Opening
Underside of
drip tray