14
Installation
1
1
Installation
The Waters
®
996 Photodiode Array (PDA) Detector operates in any standard laboratory
environment. The detector requires electrical power, sample and waste fluidic lines, and
the Millennium®
32
Workstation. Optional connections on the detector rear panel allow
communication with chart recorders, data integrators, and other instruments that are not
compatible with Millennium software control.
1.1 Installation Site Requirements
Install the Waters 996 PDA Detector (
) at a site that meets the specifications
Table 1-1 Site Requirements
Factor
Specification
Ambient temperature
4 to 40° C (39 to 104° F)
Relative humidity
20 to 80 percent,
noncondensing
Bench space
Width: 11.5 in. (29 cm)
Depth: 24 in. (61 cm)
Height: 8.125 in. (22
cm)
Bench support
Capable of supporting
31.5 pounds (14.3 kg)
Clearance
At least 4 in. (10 cm) on
the back and left sides
for ventilation