5.
Remove and discard the old fuse.
6.
Insert a new, properly rated fuse into the fuse holder.
7.
Insert the fuse holder into the receptacle, and gently push until it locks into
position.
8.
Connect the power cord to the rear panel power-entry module.
9.
Power on the module.
3.6 Cleaning the exterior of the equipment
Warning:
To avoid electric shock,
• ensure that the electrical power to the equipment is interrupted;
• when cleaning the surface of the equipment, apply water to a cloth, and
then wipe the instrument or device. Do not spray or otherwise apply
water directly onto any equipment surface.
Warning:
To avoid personal injury, use eye and hand protection during the
cleaning process.
Requirement:
Use eye protection when performing this procedure.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when
performing this procedure.