1.3.4 Autozero
You can eliminate the eluent’s contribution to conductivity using the Autozero
function. When you select Autozero, the detector measures the eluent conductivity
and assigns it a value of zero. Thus, any sample ions appear as positive or negative
measurements, relative to a zero baseline.
Figure 1–6: Absolute and relative (autozeroed) conductivity traces
December 7, 2015, 715004730 Rev. A