One of the advantages of the PoolLab 2.0 is that the (3-fold) measuring cuvette is permanently installed, but can also be changed.
It is advisable to change the cuvette (spare part) if it is scratched, dirty or generally damaged in the area of the beam path (LED sensor).
Changing the cuvette is child's play:
1)
Make sure that the unit, the cuvette and the cuvette holder are clean.
2)
Remove the light protection cover
3)
Push the cuvette holder towards you (unlock).
4)
Remove the cuvette holder
5)
Remove the old cuvette
6)
Dispose of the old cuvette in the normal household waste.
7)
Insert the new cuvette. Make sure that it is clean as a whistle so that the LED beam cannot be deflected.
8)
Put on the cuvette holder
9)
Push the cuvette holder backwards (snap it into place)
10) Put on the light protection cover and carry out a calibration (menu)
PoolLab 2.0
(cuvette change)