Maintenance
49
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
5
Inspect the gasket in the
analyzer bottle holder
assembly and replace if
necessary.
Make sure that the gasket and
bottle holder assembly mating
surfaces are free of resin
beads.
6
Install the new bottle of resin
by screwing the bottle into
place.
Tighten about ¼ turn after the
top of the bottle contacts the
gasket.
Note: If it leaks slightly, it can
be tightened further until it
stops leaking.
7
Turn on sample flow to the
analyzer and wait for
approximately one hour for the
Cation Conductivity and
Degassed Cation Conductivity
readings to drop to the
expected value.