Appendix A: USB Data Logger Extraction (Optional)
52
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
9
APPENDIX A: USB DATA LOGGER EXTRACTION (OPTIONAL)
Data is transferred between the transmitter and a USB flash drive via the memory manager. To
open the memory manager, close all windows and insert the USB flash drive into the USB host
interface. The memory manager will then open automatically. You require the relevant user rights
for access to the menu options "USB->Configure devices and "Software update". The factory
settings authorize the "Master" and "Service" users for this.
9.1
MEMORY MANAGER MENU ITEMS:
Safely remove hardware:
To prevent hardware damage or loss of data, it is necessary to
select this menu item before removing and inserted USB flash drive. Please follow the
instructions on the device's display.
Recorder Update -> USB:
Use this function for
regular retrieval of recorder data
and
continuous archiving
of measured-data histories. Measurement data that have not yet been
retrieved are stored on the flash drive together with their configuration data. The
measurement data are stored in DAT files and the configuration data in SET files. These files
can be opened and evaluated with the aid of the evaluation software. Data that has been read
out is marked internally as retrieved and the available memory display is reset to 100 %.
CAUTION:
Ensure that recorder updates are performed in a timely manner! When the ring
memory is full (display of remaining memory in device reads 0 %), measurement data in the ring
memory is lost (starting with the oldest).
Recorder Backup -> USB:
This function is used to
back up
the recorder data to
prevent
data loss
. All measurement data in the ring memory (also data already retrieved) are
transmitted to the flash memory together with their configuration data. The measurement data
are stored in DAT files and the configuration data in SET files. These files can be opened and
evaluated with the aid of the evaluation software. In contrast to Recorder Update, there is no
internal marking of recorder data and no resetting of the available memory display.
NOTE:
The "Recorder Update" and "Recorder Backup" functions are available only if the extra
code "Recording" is enabled.