Maintenance
47
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
7
MAINTENANCE
7.1
SCHEDULED MAINTENANCE
Basic maintenance of the Waltron 9096 Analyzer requires that the cation resin bottle and soda
lime cartridge be replaced periodically. In addition, the user should perform a regular 'visual
overall-check' of the wet section for immediate corrective measures, e.g. in case of leaks etc.
Cleaning of the analyzer components is best performed using a soft, non-aggressive cleaner.
The use of a logbook for cataloging consumable replacement, corrective measures and scheduled
maintenance is strongly recommended.
During the performance of the basic maintenance work, as described in this chapter, the analyzer
can not be operational. Prior to the maintenance work, all necessary precautions regarding
personal safety (protective clothing, safety glasses etc.) are to be taken into consideration.
Always be sure to label and rinse all connected tubing with water prior to removal.
List of scheduled maintenance:
Visual checks
Visually check for leaks in the system
Visually check status of the cation resin and soda lime by monitoring color change
Monthly
Replace cation resin bottle (see section 7.2)
Replace soda lime cartridge (see section 7.3). The soda lime should change from white
to a light purple color as it is consumed.