![VXI VT1422A Скачать руководство пользователя страница 167](http://html1.mh-extra.com/html/vxi/vt1422a/vt1422a_user-manual_1057857167.webp)
Advanced Programming with the VT1529B 165
Chapter 5
Field Wiring for
Excitation
Measurements
To take advantage of the on-the-fly excitation measurement feature means
that at least one channel of each VT1529B will need to dedicated to the
excitation voltage measurement. Any of the 32 channels can be used for this
purpose. If the application requires more frequent measurements than once
per scan, multiple channels can be dedicated and a separate excitation
measurement channel for each strain measurement channel can even be
used. The excitation measurement channel must be specified for each strain
measurement channel using SENS:FUNC:STR:
<bridge_type>
:POST.
For the dedicated excitation measurement channel, simply leave the RJ-45
ports for the channel unconnected. (Putting a short between the strain gage
Sense + and Sense - inputs, however, may be desired. See “Settling Time
Considerations” on page 178 for a discussion of why this may be desired.)
This works because there is an internal 10 k
Ω
resistor between the Bridge
Excitation input on the front of the VT1529B and the Excitation Sense + and
- inputs for each channel in the 8-channel bank that each Bridge Excitation
input provides power to. Thus, making an excitation voltage measurement
on an unconnected channel input will measure the voltage supplied to the
Bridge Excitation input.
Note
For systems with multiple VT1529Bs, the excitation measurements must be
made on the same VT1529B as the strain gage channels that use that
excitation voltage. This means that for a system with two VT1529Bs, the
excitation voltage cannot be measured on only one VT1529B with that
measurement being applied to the strain gage measurements on both
VT1529Bs.
Note
The wiring for the strain gage channels is unchanged.
Strain Measurement
Command
Sequence
The steps for making strain measurements with the new VT1529B
functionality may already be familiar. The unstrained voltage and gage
factor for the strain channels need to be provided. The excitation channel for
each of the strain channels must also specified as well as the channels to be
scanned. One difference is that the excitation channel(s) must be included in
the scan list. Figure 5-1 shows the sequence of VXI
plug&play
driver calls
to configure and scan strain channels using the new functionality.
Содержание VT1422A
Страница 2: ...This page is blank...
Страница 8: ...8 Notes...
Страница 22: ...22 Support...
Страница 57: ...Field Wiring 57 Chapter 2...
Страница 58: ...58 Field Wiring Chapter 2...
Страница 96: ...94 Programming the VT1422A VT1529A B for Remote Strain Measurement Chapter 3...
Страница 162: ...160 Programming the VT1422A for Data Acquisition and Control Chapter 4...
Страница 224: ...222 Creating and Running Algorithms Chapter 6...
Страница 419: ...VT1422A Command Reference 417 Chapter 7 Notes...
Страница 420: ...418 VT1422A Command Reference Chapter 7 Notes...
Страница 425: ...Specifications 423 Appendix A Thermocouple Type E 200 800 C SCPs VT1501 02 03A...
Страница 426: ...424 Specifications Appendix A Thermocouple Type E 200 800 C SCPs VT1508 09A...
Страница 427: ...Specifications 425 Appendix A Thermocouple Type E 0 800 C SCPs VT1501A 02A 03A...
Страница 428: ...426 Specifications Appendix A Thermocouple Type E 0 800 C SCPs VT1508A 09A...
Страница 429: ...Specifications 427 Appendix A Thermocouple Type E Extended SCPs VT1501A 02A 03A...
Страница 430: ...428 Specifications Appendix A Thermocouple Type E Extended SCPs VT1508A 09A...
Страница 431: ...Specifications 429 Appendix A Thermocouple Type J SCPs VT1501A 02A 03A...
Страница 432: ...430 Specifications Appendix A Thermocouple Type J SCPs VT1508A 09A...
Страница 433: ...Specifications 431 Appendix A Thermocouple Type K SCPs VT1501A 02A 03A...
Страница 434: ...432 Specifications Appendix A Thermocouple Type R SCPs VT1501A 02A 03A...
Страница 435: ...Specifications 433 Appendix A Thermocouple Type R SCPs VT1508A 09A...
Страница 436: ...434 Specifications Appendix A Thermocouple Type S SCPs VT1501A 02A 03A...
Страница 437: ...Specifications 435 Appendix A Thermocouple Type S SCPs VT1508A 09A...
Страница 438: ...436 Specifications Appendix A Thermocouple Type T SCPs VT1501A 02A 03A...
Страница 439: ...Specifications 437 Appendix A Thermocouple Type T SCPs VT1508A 09A...
Страница 440: ...438 Specifications Appendix A 5k Thermistor Reference SCPs VT1501A 02A 03A...
Страница 441: ...Specifications 439 Appendix A 5k Thermistor Reference SCPs VT1508A 09A...
Страница 442: ...440 Specifications Appendix A RTD Reference SCPs VT1501A 02A 03A...
Страница 443: ...Specifications 441 Appendix A RTD SCPs VT1501A 02A 03A...
Страница 444: ...442 Specifications Appendix A RTD SCPs VT1508A 09A...
Страница 445: ...Specifications 443 Appendix A 2250 Thermistor SCPs VT1501A 02A 03A...
Страница 446: ...444 Specifications Appendix A 2250 Thermistor SCPs VT1508A 09A...
Страница 447: ...Specifications 445 Appendix A 5k Thermistor SCPs VT1501A 02A 03A...
Страница 448: ...446 Specifications Appendix A 5k Thermistor SCPs VT1508A 09A...
Страница 449: ...Specifications 447 Appendix A 10k Thermistor SCPs VT1501A 02A 03A...
Страница 450: ...448 Specifications Appendix A 10k Thermistor SCPs VT1508A 09A...
Страница 454: ...452 Specifications Appendix A...
Страница 480: ...478 VT1529A B Verification Calibration Appendix C...
Страница 484: ...482 Glossary Appendix D Notes...