148 Programming the VT1422A for Data Acquisition and Control
Chapter 4
Reading Status
Groups Directly
It may be desirable to directly poll status groups for instrument status rather
than poll the Status Byte for summary information.
Reading Event Registers
The Questionable Data, Operation Status and Standard Event Groups all
have Event Registers. These Registers log the occurrence of even temporary
status conditions. When read, these registers return the sum of the decimal
values for the condition bits set, then are cleared to make them ready to log
further events. The commands to read these Event Registers are:
STAT:QUES:EVENT?
Questionable Data Group Event
Register
STAT:OPER:EVENT?
Operation Status Group Event
Register
*ESR?
Standard Event Group Event
Register
Clearing Event Registers
To clear the Event Registers without reading them, execute:
*CLS
clears all group’s Event
Registers
Reading Condition
Registers
The Questionable Data and Operation Status Groups each have a Condition
Register. The Condition Register reflects the group’s status condition in
"real-time." These registers are not latched so transient events may be
missed when the register is read. The commands to read these registers are:
STAT:QUES:COND?
Questionable Data Group
Condition Register
STAT:OPER:COND?
Operation Status Group
Condition Register
VT1422A Background Operation
The VT1422A inherently runs its algorithms and calibrations in the
background mode with no interaction required from the driver. All resources
needed to run the measurements are controlled by the on-board Control
Processor (DSP).
The driver is required to setup the type of measurement to be run, modify
algorithm variables and to unload data from the card after it appears in the
CVT or FIFO. Once the INIT[:IMM] command is given, the VT1422A is
initiated and all functions of the trigger system, measurement scanning,
and algorithm execution are controlled by its on-board control processor.
The driver returns to waiting for user commands. No interrupts are required
for the VT1422A to complete its measurement.
While the module is scanning and/or running algorithms, the driver can
be queried for its status and data can be read from the FIFO and CVT.
The ABORT command may be given to force continuous execution to
complete. Any changes to the measurement setup will not be allowed until
the TRIG:COUNT is reached or an ABORT command is given. Of course,
any commands or queries can be given to other instruments while the
VT1422A is running algorithms.
Содержание VT1422A
Страница 2: ...This page is blank...
Страница 8: ...8 Notes...
Страница 22: ...22 Support...
Страница 57: ...Field Wiring 57 Chapter 2...
Страница 58: ...58 Field Wiring Chapter 2...
Страница 96: ...94 Programming the VT1422A VT1529A B for Remote Strain Measurement Chapter 3...
Страница 162: ...160 Programming the VT1422A for Data Acquisition and Control Chapter 4...
Страница 224: ...222 Creating and Running Algorithms Chapter 6...
Страница 419: ...VT1422A Command Reference 417 Chapter 7 Notes...
Страница 420: ...418 VT1422A Command Reference Chapter 7 Notes...
Страница 425: ...Specifications 423 Appendix A Thermocouple Type E 200 800 C SCPs VT1501 02 03A...
Страница 426: ...424 Specifications Appendix A Thermocouple Type E 200 800 C SCPs VT1508 09A...
Страница 427: ...Specifications 425 Appendix A Thermocouple Type E 0 800 C SCPs VT1501A 02A 03A...
Страница 428: ...426 Specifications Appendix A Thermocouple Type E 0 800 C SCPs VT1508A 09A...
Страница 429: ...Specifications 427 Appendix A Thermocouple Type E Extended SCPs VT1501A 02A 03A...
Страница 430: ...428 Specifications Appendix A Thermocouple Type E Extended SCPs VT1508A 09A...
Страница 431: ...Specifications 429 Appendix A Thermocouple Type J SCPs VT1501A 02A 03A...
Страница 432: ...430 Specifications Appendix A Thermocouple Type J SCPs VT1508A 09A...
Страница 433: ...Specifications 431 Appendix A Thermocouple Type K SCPs VT1501A 02A 03A...
Страница 434: ...432 Specifications Appendix A Thermocouple Type R SCPs VT1501A 02A 03A...
Страница 435: ...Specifications 433 Appendix A Thermocouple Type R SCPs VT1508A 09A...
Страница 436: ...434 Specifications Appendix A Thermocouple Type S SCPs VT1501A 02A 03A...
Страница 437: ...Specifications 435 Appendix A Thermocouple Type S SCPs VT1508A 09A...
Страница 438: ...436 Specifications Appendix A Thermocouple Type T SCPs VT1501A 02A 03A...
Страница 439: ...Specifications 437 Appendix A Thermocouple Type T SCPs VT1508A 09A...
Страница 440: ...438 Specifications Appendix A 5k Thermistor Reference SCPs VT1501A 02A 03A...
Страница 441: ...Specifications 439 Appendix A 5k Thermistor Reference SCPs VT1508A 09A...
Страница 442: ...440 Specifications Appendix A RTD Reference SCPs VT1501A 02A 03A...
Страница 443: ...Specifications 441 Appendix A RTD SCPs VT1501A 02A 03A...
Страница 444: ...442 Specifications Appendix A RTD SCPs VT1508A 09A...
Страница 445: ...Specifications 443 Appendix A 2250 Thermistor SCPs VT1501A 02A 03A...
Страница 446: ...444 Specifications Appendix A 2250 Thermistor SCPs VT1508A 09A...
Страница 447: ...Specifications 445 Appendix A 5k Thermistor SCPs VT1501A 02A 03A...
Страница 448: ...446 Specifications Appendix A 5k Thermistor SCPs VT1508A 09A...
Страница 449: ...Specifications 447 Appendix A 10k Thermistor SCPs VT1501A 02A 03A...
Страница 450: ...448 Specifications Appendix A 10k Thermistor SCPs VT1508A 09A...
Страница 454: ...452 Specifications Appendix A...
Страница 480: ...478 VT1529A B Verification Calibration Appendix C...
Страница 484: ...482 Glossary Appendix D Notes...