Front Panel LED Indications for Message-Based Device Operation
The GPIB-VXI indicates the progress of its self-test with the FAILED, TEST, and ONLINE
LEDs. The LED indications are shown in Table 2-11. A successful system startup sequences
through the first five states. The probable cause of failure shown for each combination of LEDs
is valid if the FAILED LED is lit five seconds following system startup. The LED indications
are identical for Non-Slot 0 Message-Based device and Slot 0 Message-Based device operation.
Table 2-11. Front Panel LED Indications for Message-Based Device Operation
Sequence
FAILED TEST
ONLINE
State
Point of Failure
1
OFF OFF OFF No
power
2
ON
OFF
OFF
In self-initialization Failed before self-test
3
ON ON OFF In
self-test
Failed
self-test
4
OFF ON ON Waiting
for
BNO
5
OFF OFF ON Online
ON OFF ON Failed
Failed
while
online
Slot 0 Message-Based Device Configuration
The GPIB-VXI is configured for Slot 0 Message-Based device operation by enabling the
VXIbus Slot 0 functions, setting the model code to 0FFh, and setting the logical address to a