Diagnostic Tests
Chapter 5
GPIB-VXI User Manual
5-4
© National Instruments Corporation
Table 5-2. Diagnostics Mode Menu Option Descriptions
Selection
Description
Default
Runs the selected subset of tests (1-9).
Stepping Steps
Runs the selected subset of steps (1-126).
Looping Tests
Runs the selected subset of tests (1-9) repeatedly.
Looping Steps
Runs the selected subset of steps (1-126) repeatedly.
Slow Tests
Runs the selected subset of tests (1-9) with a pause
between the execution of each test.
Slow Steps
Runs the selected subset of steps (1-126) with a pause
between the execution of each step.
Slow Looping Tests
Runs the selected subset of tests (1-9) repeatedly with a
pause between the execution of each test.
Slow Looping Steps
Runs the selected subset of steps (1-9) repeatedly with a
pause between the execution of each step.
Display Step Line
Runs the selected subset of steps (1-126) and displays the
subroutine commands executed for each step.
Display/Hold Lines
Runs the selected subset of steps (1-126) and displays the
subroutine commands executed for each step. With this
mode you must input a keystroke to continue after each
command is executed.
Over Night Loop
Runs the selected subset of steps (1-126) repeatedly until
an error is encountered or the module is reset. If you select
the
All
option from the second-level menu, all steps but
the DIP switch and trigger steps are executed. Since the
DIP switch and trigger steps are interactive, they are not
included in the Over Night Loop tests.
By default, test result messages are printed to the serial port as each test is executed, and
diagnostics execution terminates if an error occurs. Using the Diagnostics Mode menu, you can
suppress test result message printing and the abort-on-error logic. Suppress printing with the
p
toggle selection, and the abort-on-error logic with the
e
toggle selection.
You can recall the last error at any time (before the module is powered-down or restarted) with
the
r
selection. The last detected error is retained even if the diagnostics mode is exited and re-
entered. This is a convenient feature that you can use under these conditions: