Operating Instructions | symphony 5.3 A – 06/2009
4
Description of the device
4
Description of the device
25
4.8 90 °C decontamination routine (optional)
Certified decontamination routine
The optional 90 °C decontamination routine is used to decontaminate the
complete work space including all installed components and sensors.
During this routine, a moist and wet atmosphere with highly decontamina-
ting effect is created for 9 hours at a temperature of 90° C.
The effectiveness of the decontamination routine has been tested and cer-
tified by independent institutes. Information about these tests is available
at request from VWR.
The entire program run of the decontamination routine will take approx 25
hours.
After the run has been completed, the device must be reactivated using the
autostart routine.
NOTE!
Thermal protection:
Restart in cause of failure
If the thermal protection for the device responds, the 90 °C
decontamination routine can only be be started after the cause
of the failure has been repaired or reset.