REFRIGERATOR OPERATION

 

After the unit is properly installed and power is applied, it will take some time before the 
system is cooled down to temperature and cycling normally. You should wait 8 hours on the 
first startup before beginning to add product to the unit. This ensures that the unit is installed 
and operating properly before being put into service. On subsequent startups (after cleanings, 
for example), this wait time can be reduced to about 3 hours.   After this wait time, the unit 
should be cycling in the design temperature range. The units are calibrated before leaving the 
factory, so no adjustment should be necessary.  

Loading the units will again cause temperature to rise as the warmer product is introduced into 
the compartment. If a large amount of product is to be introduced, it is a good idea to do it in 
stages, allowing several hours between stages to allow temperature to stabilize again before 
introducing additional warm product. This will minimize the temperature transient while 
loading. Other tips for successful loading include:  

-

 

Leave about 2-3 inches of free space along the back and sides of the unit to allow for proper 
air flow and, therefore, more even cooling of the product.  

-

 

When loading the top shelf, avoid blocking the evaporator fan(s) (if installed). There should 
be at least 4 inches of clearance below the fan(s) to allow proper air flow.  

-

 

Do not overload the unit. A full load is about 70 to 75% full. Additionally, the load should be 
distributed evenly from top to bottom and side to side for best results.  

-

 

Minimize the time the door is open. On top of letting the cold air out, you are also letting 
warm, moist air in which can result in more condensation and/or frost in the unit.  

 

Model SCLP-3004 will go into a defrost cycle every 6 hours to remove frost build-up around the 
evaporator coils. This cycle can last up to 25 minutes. During this time the compressor is turned 
off and defrost heaters are turned on in the evaporator section. This will cause a rise in air 
temperature inside the unit but should not significantly affect product temperatures.  

Remember that the units are calibrated to the design temperature range before leaving the 
factory. We also do extensive testing to ensure that these temperatures will result in product 
temperatures in the desired band. There should be no need to adjust the temperature control 
on these units, but if it is necessary, they can be adjusted using the manual thermostat on the 
back of the unit

.  

It is STRONGLY RECOMMENDED that you contact our Technical Support Department prior to 
performing any temperature adjustments to ensure the adjustment is necessary and, if so, it 
is performed correctly.

  

Содержание SCLP-2004

Страница 1: ... Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is being unpacked When concealed damage is discovered stop unpacking immediately and contact the carrier immediately to report the claim and request an inspection This should be done as soon as possible and in any case must be done wit...

Страница 2: ...AL SERVICE please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left upper wall of the unit For convenience you may want to record the following information here for easy access in the future Model number __________________________________ Serial number ___________________...

Страница 3: ...ry receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along with a formal request for an inspection report Follow up with a phone call Their contact information can be found on the carrier s web site The second type of shipping damage is concealed damage This t...

Страница 4: ...re installed This is acceptable and recommended on this unit POWER SOURCE The supply circuit to this cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection VOLTAGE SHOULD NOT VARY MORE THAN 5 FROM SERIAL PLATE RATINGS A separate circuit is recommended to prevent possible loss of pro...

Страница 5: ...elves Wire shelves are oriented so that cross support bars are facing down Place shelves on shelf clips making sure all corners are seated properly TEMPERATURE CONTROL NOTE The controls are preset at the factory to provide the desired air and product temperatures inside of the units and require no further adjustment Please contact the manufacturer s Technical Support Department before making any a...

Страница 6: ...top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inches of clearance below the fan s to allow proper air flow Do not overload the unit A full load is about 70 to 75 full Additionally the load should be distributed evenly from top to bottom and side to side for best results Minimize the time the door is open On top of letting the cold air out you are also lettin...

Страница 7: ...a battery backup which will maintain the alarm system for several hours after a power failure to allow for continued alarm and notification functions The temperature monitor system does NOT control the temperature of the unit That function is performed by the Microprocessor Digital Temperature Control CONTROL CENTER PANEL The LCD Display shows the temperature in ºC In the event of a loss of power ...

Страница 8: ... value The DOWN button has a different function for each operating mode In Normal Run mode pressing the button causes the low temperature set point to be displayed In Programming mode pressing the button decreases the displayed parameter value BASIC OPERATION The Control Center Monitor has two modes NORMAL RUN MODE and PROGRAMMING MODE This section describes the unit in Normal Run Mode In Normal R...

Страница 9: ... To test the battery push the ALARM TEST button on the front panel The following should happen The ALARM LED will flash red The beeper alarm will begin to chirp The remote alarm relay will switch PROGRAMMING To program The Control Center Monitor 1 Hold the MUTE button for two seconds to bring the unit into Programming mode 2 The display will show the first parameter that can be programmed high tem...

Страница 10: ... delay This alarm delay time will prevent transient nuisance alarms The alarm delay time can be set for 0 to 30 minutes rd Relay Delay If there is an alarm condition the relay contacts will switch The relay delay parameter allows the user to set the amount of time delay from the start of the alarm condition to the time the relay contact switches This keeps the relay from switching during short nui...

Страница 11: ...ls remote alarms or central monitoring systems that an alarm condition has occurred at the unit being monitored The relay will pull in when line power is present If an alarm condition lasts for a designated time the relay delay the relay will drop out causing the relay contacts to switch If the alarm condition does not last long enough for the relay delay to time out the relay will not go into its...

Страница 12: ...ng press the SET key once to access the Fahrenheit Celsius mode The display shows the current status F for degrees Fahrenheit or C for degrees Celsius Press either up or down arrow to toggle between the F or C F or C Fahrenheit or Celsius Scale Step 2 Press SET key again to access the setpoint mode The LCD will display the current setpoint and S1 will be blinking to indicate that the control is in...

Страница 13: ...correct if necessary EP Appears when the probe is open shorted or sensing a temperature that is out of range Check to see if the sensed temperature is out of range If not check for probe damage by comparing it to a known ambient temperature between 30 F and 220 F Replace the probe if necessary EE Appears if the EPROM data has been corrupted This condition cannot be field repaired Replace the contr...

Страница 14: ...eplacements are pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or side walls Evaporator is blocked by frost or ice Remove the products unplug the refrigerator or freezer power and allow the unit to defrost If the problem still exists call for service 3...

Страница 15: ...o prevent leakage Door will not close The unit is not level Refer to the Leveling section at the beginning of this manual Check for dirt and debris or items passing through the door seal MOISTURE DURING THE SUMMER SEASON The amount of moisture condensation or high humidity related issues increase during the summer and in most cases will self resolve when the weather cools down Please note a refrig...

Страница 16: ...tablespoons of baking soda in 1 quart of warm water Be sure to wring excess water out of sponge or cloth before cleaning around controls light bulb or any electrical parts Door Gaskets Soap and water Wipe gaskets and their seating surfaces with a clean soft cloth Shelves Soap and water Do not wash removable shelves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not...

Страница 17: ...TY In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compressors to be free from defects in both material and workmanship under normal use and service for a period of three 3 additional years from the end of the initial two 2 year warranty period but not to exceed five 5 years after shipment from Horizon Scientific Inc...

Страница 18: ...ood or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with the printed instructions provided or damage caused by the storage of any corrosive material that comes in contact with the interior or exterior portions of the cabinet or the use of spark producing equipment or containers such as galvanized or carb...

Отзывы: