Part number 20047, Rev 0

 

13 

 

rd (Relay Delay)

 

If there is an alarm condition, the relay contacts will switch.  The relay delay parameter allows 
the user to set the amount of time delay from the start of the alarm condition to the time the 
relay  contact  switches.    This  keeps  the  relay  from  switching  during  short  “nuisance  alarms”.  
The operator may set the relay delay time for 0 to 30 minutes.

 

SIL (Alarm Silence, or Alarm Mute Time) 

This  parameter  sets  the  number  of  minutes  the  beeper  will  “chirp”  when  the  mute  button  is 
pressed during an alarm condition.  After the Silence Time elapses, if the alarm condition still 
exists, the beeper will begin to sound at full volume again.  This parameter can be set from 5 to 
120 minutes. 

Remote Alarm Contacts 

The  Remote  Alarm  Relay  is  a  single-pole  double-throw  (SPDT)  relay  that  changes  state 
(switches) with the loss of line power or with the presence of an alarm condition. 
NOTE: The relay is rated for pilot duty operation only. Do NOT use it for control switching. The 
relay contacts and connections are rated at 30v/1A. 
In normal applications, the relay signals remote alarms or central monitoring systems that an 
alarm condition has occurred at the unit being monitored. 
The relay will “pull-in” when line power is present. If an alarm condition lasts for a designated 
time  (the “relay delay”),  the  relay  will  “drop  out”,  causing the  relay  contacts  to  switch.  If the 
alarm condition does not last long enough for the relay delay to time out, the relay will not go 
into its alarm position. The relay will also “drop out” if  power to the alarm is lost, acting as a 
power loss detector.  
The diagram below is an example of hooking up our external relay output to a remote alarm. 

 
 

Содержание SCCP-19

Страница 1: ...agent driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is...

Страница 2: ...MANUFACTURER S TECHNICAL SERVICE DEPARTMENT please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can b...

Страница 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Страница 4: ...t needed PLEASE NOTE Your refrigerator or freezer is designed as INDOOR USE and should be operated at air conditioned room temperature between 65 F to 85 F Remove the lower grill assembly from the fro...

Страница 5: ...the plastic door stoppers see Image 4 5 and 6 CASTERS INSTALLATION It is important to ensure the casters are installed properly and securely in order to obtain the maximum strength and stability of th...

Страница 6: ...on cord Be sure your unit is properly grounded Use the 3 prong plug provided into a 3 prong grounded outlet Only this method complies with national electrical codes local codes and ordinances Unless t...

Страница 7: ...nd product temperatures inside of the units and require no further adjustment Please contact the manufacturer s Technical Support Department before making any adjustment to make sure the adjustment is...

Страница 8: ...king the evaporator fan s if installed There should be at least 4 inches of clearance below the fan s to allow proper air flow Do not overload the unit Maximum suggested load is about 75 of chamber ca...

Страница 9: ...ated with the alarm system i e delay times system status a battery backup which will maintain the alarm system for several hours after a power failure to allow for continued alarm and notification fun...

Страница 10: ...pressed The MUTE PROGRAM button has three functions In Normal Run mode if the unit is beeping pressing the button will mute the beeper for a preset time Each time the button is pressed the mute time r...

Страница 11: ...the BEEPER and ALARM LED will both activate Press the MUTE PROGRAM button to stop the beeper for a specified time changing the sound to short chirps When the temperature goes back within range and the...

Страница 12: ...n a short period 5 Once the Monitor is in Programming mode each time the Mute button is pressed the display steps to the next parameter As the display steps to the next parameter the value of the prev...

Страница 13: ...mote Alarm Relay is a single pole double throw SPDT relay that changes state switches with the loss of line power or with the presence of an alarm condition NOTE The relay is rated for pilot duty oper...

Страница 14: ...of the unit A touch keypad allows the user to easily select the display units set point differential set point and heating or cooling mode Please Note The digital temperature controller has been fact...

Страница 15: ...et point and S1 will be blinking Press the up and down keys to adjust to the desired setting S1 blinking Setpoint Temperature Press SET key again to access the differential mode The LCD will display t...

Страница 16: ...y EP Appears when the probe is open shorted or sensing a temperature that is out of range Check to see if the sensed temperature is out of range If not check for probe damage by comparing it to a know...

Страница 17: ...placements are pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerat...

Страница 18: ...prevent leakage Door will not close The unit is not level Refer to the Leveling section at the beginning of this manual Check for dirt and debris or items passing through the door seal MOISTURE DURIN...

Страница 19: ...sehold cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smooth handles Do not use a dry cloth to clean smooth handles For Glass Door Units clean the glass with a mild detergent a...

Страница 20: ...th pressurized air CO2 or nitrogen REMEMBER THE CLEANING OF THE CONDENSER IS NOT COVERED BY THE WARRANTY AND IS YOUR RESPONSIBILITY ANY DAMAGE CAUSED BY FAILURE TO KEEP THE CONDENSER CLEAN IS ALSO NOT...

Страница 21: ...Y In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compressors to be free from defects in both material and workmanship...

Страница 22: ...od or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with the printed instructions provided or damage cause...

Отзывы: