Maintenance |
Chapter 10
NOTE!
Excessive CO
2
content:
Excessive CO
2
content after the calibration can be reduced by leaving the device doors open for
approx. 30 seconds!
CO
2
calibration procedure
1. Activate the calibration:
Keep the
key depressed for 5 seconds.
All operating panel indicators flash.
2. Indicate the nominal value:
Press the
key.
The set nominal value of 5% is displayed.
3. Enter the measured value (destination value):
Press the
+
keys.
or
Press the
+
keys.
Destination value e.g. 5.6%.
4. Accept the destination value:
Press the
key.
The CO
2
display momentarily shows “CAL”,
then the corrected actual value (measured destination value 5.6%) is displayed.
5. Cancel the calibration process:
Press any key.
The temperature display and the CO
2
display show the actual values.
Содержание CO2 Incubator Basic
Страница 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Страница 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Страница 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Страница 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Страница 77: ...VWR CO2 Incubator Basic 75 Technical data Chapter 12 Average gas consumption Figure 24 CO2 consumption ...
Страница 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Страница 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Страница 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Страница 85: ......
Страница 86: ......