VWR
CO2 Incubator Basic |
45
Handling and control |
Chapter 7
Activating the auto-start routine
The auto-start function is an automated routine for the start and the subsequent adjustment of the
CO
2
measuring system. After the start, the device control adjusts the temperature to the set
nominal value while humidity is generated. When temperature and relative humidity have reached
constant values, the CO
2
measuring system is automatically adjusted to these values, and the
work space is supplied with the preset quantity of CO
2
.
NOTE!
Application of the routine:
To ensure that the specified accuracy of the CO
2
measuring system is maintained, the device
should always be started using the auto-start routine after the nominal temperature setting has
been changed by more than 1 °C or after extended interruptions of the operation of the device. The
auto-start routine should be run at least every three months on the occasion of cleaning and
maintenance works.
Running the routine usually takes 5 to 7 hours. At low room temperatures and when the device is
cold, it may take up to 10 hours until the auto-start routine has been completed. If the glass door
is opened or if the power supply of the device is interrupted while the routine is running, the routine
is interrupted and rerun after the glass door has been closed and after the power supply has been
reestablished.
At the start of the auto-start routine, the work space atmosphere must consist only of ambient air.
The floorpan must be filled with a sufficient quantity of water!
1. Open both doors until the acoustic alarm sounds after 30 seconds:
All current actual values flash on the displays, the “door” LED illuminates, after 30
seconds the acoustic alarm sounds.
2. Enter nominal values:
See Sections “Setting the nominal temperature value” and “Setting the CO2 nominal value”.
3. Activate the auto-start routine:
Keep the key
depressed for 5 seconds.
The auto-start LED flashes.
4. Close all device doors:
The temperature display shows the actual value, the CO
2
display shows “run”, the “door”
LED goes off.
Содержание CO2 Incubator Basic
Страница 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Страница 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Страница 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Страница 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Страница 77: ...VWR CO2 Incubator Basic 75 Technical data Chapter 12 Average gas consumption Figure 24 CO2 consumption ...
Страница 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Страница 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Страница 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Страница 85: ......
Страница 86: ......