CO2 Incubator Basic |
59
Cleaning and disinfection |
Chapter 9
4. Complete the routine:
Keep the
key depressed for 5 seconds.
The display returns to the normal operating state (incubation operation).
Cancelling the 90 °C decontamination routine
1. Cancel a routine phase:
Keep the
key depressed for 5 seconds.
The routine advances to the next phase.
2. Cancel the routine completely:
To completely cancel the routine, each phase must be skipped individually by pressing the
key until a remaining run time of 0 hours is displayed.
When the last phase (postheating phase) has been skipped, the routine is completely
cancelled.