![VWR 630-3072 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/vwr/630-3072/630-3072_instruction-manual_1057554010.webp)
10
9.6
Use of additional lens (SZB260/SZT260)
Screw the desired additional lens on the microscope
body. (Fig. 14)
Each additional lens has a specific Working Distance
(see table below).
The stroke of the focus adapter could not compen-
sate the different working distances of the several
additonal lenses.
Fig. 14
Total magnification used can be calculated as:
Eyepiece magnification * Zoom magnification * Objective lens magnification.
Eyepiece
10X
15X
20X
Field number
21
15
10
Objective
Total mag.
F.O.V. (mm)
Total mag.
F.O.V. (mm)
Total mag.
F.O.V. (mm)
0,5X (W.D.165 mm)
3.5X-22.5X
59.99-9.32
5.25X-33.75X
28.57-4.44
7X-45X
14.28-2.22
0.75X (W.D. 114 mm)
5.25X-33.75X
40-6.21
7.87X-50.62X
19.06-2.96
10.50X-67.5X
9.52-1.48
1X (W.D. 100 mm)
7X-45X
29.99-4.66
10.5X-67.5X
10.95-2.22
14X-90X
7.14-1.11
1.5X (W.D. 47 mm)
10.50X-67.5X
20-3.10
15.75X-101.25X
9.52-1.48
21X-135X
4.76-0.74
Fig. 15
Fig. 16
9.6.1 Use of 0.5X additional lens
0.5X additional lens has a Working Distance of 165
mm and therefore the total stroke of the focus pillar
cannot compensate.
In order to let the 0.5X additional lens working pro-
perly, proceed as follows:
1. Screw the additional lens on the microscope body
as already described in chapter 8.6.
2. Uninstall the microscope body from the head hol-
der.
3. Install the provided spacer
①
in the head holder
and screw the fixing knob
②.
(Fig. 15)
4. Re-install the microscope body onto the spacer
and screw the fixing knob
③.
(Fig. 16)
5. Begin working normally.
③
②
①