Instruction Manual INCU-Line Cooled
Page
45
/
50
13.1. Possible defects
Malfunction
What to check
What to do
The unit is not working
Check the voltage in the socket
Measure the voltage in the
socket; if necessary, change the
fuses in the electric installation
in the building
Check if the unit is plugged in
Plug in the unit
Check the status of a circuit-breaker on the
back of the device
Press the circuit breaker on the
back of the device.
Check if the power cable is broken
Change the cable
The unit is not cooling
down
Check if the evaporator is dirty
Clean the evaporator
Check if the unit is exposed to direct sunlight
Change the location of the unit
Check if the unit is located near a radiator
Change the location of the unit
Check if the door of the unit is closed properly
Clean the gasket
The unit is not heating
up
Check if the door of the unit is closed properly
Clean the gasket
Check if the fan is turned on
If not please contact service
Does the installation site meet the installation
conditions?
Adapt to the installation require-
ments on site (according to
point 3 chapter 1)
The unit is working too
loudly
Check if the unit is not touching other objects
(e.g. furniture etc.)
Remove other objects
Check if the doors are properly leveled
If the doors are correctly leveled
and still are sagging, please
contact service