Instruction Manual INCU-Line Cooled
Page
10
/
50
To remove a shelf, perform the above steps in reverse order. To remove the shelf holder from the slots, lift
it up and slide it towards the rear of the chamber.
3.2. Remarks on the type of samples
Water may gather on the bottom of the chamber. It is a result of condensation of the water vapor located in
air if the set temperature is considerably lower than the ambient temperature.
The amount of water depends on the following factors:
-
Differences between ambient and set temperatures
-
Number and frequency of door openings
-
Temperature of samples
Do not use any cardboard boxes, sponges and other hygroscopic materials for storing the samples since
they may increase the relative humidity in the chamber.
3.3. Remarks on the placement of samples
To provide proper air circulation and stable conditions in which the samples are stored in the chamber, it is
necessary to keep the following rules:
-
the max height of the samples should not exceed 1/3 of the space below the shelves
-
the samples should be placed in such a way that so that the horizontal surface between the
containers does not exceed 1/3 of the width and height of the empty shelf
-
the space between the samples and between the samples and the wall should be more or less equal
The picture below is an example of the placement of samples in the chamber.
If water gathers, use a dry cloth to wipe the bottom of the chamber.