Instruction Manual INCU-Line Cooled
Page
15
/
50
Figure 3 - Status - icons
Figure 4
5.3.2.1. Status
Internal light is switched on
Closed door, open door. The number above the icon presents open door
counter, press the icon to cancel the counter. The counter is also cancelled
during turning on the device
Icon for the fan. Rotating icon shows that the fan is running at 100%. It is
not possible to adjust the fan speed. Icon does not move at status program
off or when the fan is defect
Ramp status: Chamber is currently heating up / cooling down to reach set
temperature
Set temperature is reached
Icon is only visible when the chamber is cooling
Icon is visible only when the chamber increases the temperature
Icon is visible only when the defrosting program is running
5.3.2.2. Information
There are icons on information panel
– one of them opens
Alarms panel
» chapter 5.3.3.
second
Status panel
» chapter 5.3.4.