Instruction Manual INCU-Line Premium
Page
47
/
61
5.21.
Corrections
In this window
(Figure 50)
you can set correct:
The temperature indicated on the display by adding the correction value. The set correction value is
taken in the whole temperature range operation of the device. For example, if the average tempera-
ture displayed by the device indicates 37
°C and the average temperature measured by independent,
external sensor indicates 37
,5°C, the correction should be set on +0,5°C. The average temperature
should be calculated from chosen period of time e.g. 30min. The correction range of -
5°C to +5°C.
The unit has been calibrated by the manufacturer in accordance with applicable norms. The
temperature shown on the display corresponds with a great accuracy to the temperature in the
geometrical centre of the chamber. For the correct operation of the unit it is not necessary to use
User’s calibration.
The user is performing temperature correction on his own responsibility and he must be aware of
consequenc
es of changing of manufacturer’s settings. If the equipment was calibrated, calibra-
tion certificate loses its validity.
Figure 49 - Defrosting