Instruction Manual INCU-Line Premium
Page
22
/
61
With more parameters you can scroll the panel up and down
5.6.2. Segments edition
Press the button
, and the first program segment will appear
(Figure 13, Figure 14, Figure 15)
.
In this window you can set:
temperature
– target temperature which the device is to achieve in this segment, (can’t be lower
than under temperature prot2°C and higher than over temperature protection -2°C).
time
– the time of maintaining the set temperature ([d hh:mm]) (days, hours and minutes) it is possi-
ble to select continuous work in the last segment.
ramp time
– the time of reaching the set temperature ([d hh:mm]) (days, hours, minutes)
fan
– fan efficiency in percent,
ramp fan
– fan efficiency during reaching the set temperature
The active value is highlighted in blue.
The item highlighted in red means that the value is out of range and you should enter another one, eg the
temperature is above / below the operating range of the device or the protection temperature.
Figure 13 - Program segment edition