Instruction Manual INCU-Line Premium
Page
7
/
61
The place of location of the unit should meet the following conditions:
Ambient temperature b10
°C...+28°C
Ambient relative humidity of 60%
Low dust environment
Appropriately ventilated according to its size
Placed on a hard and stable surface
Placed at least 100 mm away from the wall
Height of the room must be at least 300 mm greater than the height of the unit
Keep away from direct sunlight
Kept away from any heat sources
Not designed to be built-in e.g. within a wall cavity
Placed near a mains socket 230V/50Hz.
Non-compliance with the above recommendations may result in damage to the unit or impair tech-
nical parameters.
The electric installation should meet the following conditions:
The electric installation should be secured by a 16 A antisurge fuse.
The IL 240CR, IL 400CR and IL 750CR units are
equipped with leveling castors. After the unit has been
placed in its destination, lock the castors and level the
unit using the red screw located inside t
he castor’s cas-
ing. You can adjust the level using fingers or a flat wrench
size 13.
Do not use the castors for transporting the unit! The castors should be used only to move device
in the lab.
The device is powered by AC 220-240V/50Hz. You must be connected to an earthed
mains socket to avoid electric shock in case of potential malfunction.