Variable
value
Operat
ing
mode
Characteristics
Heater
Fan
Target temperature
0
Air
supply
Off
Depedning on value of
MB_ADDR_FAN_SPEED
n/a
1
Heating
1
ALG
one output is set to the
heater ( terminal block
J1-H)
ALG
MB ADDR ZONE TEMPER
ATURE_TARGET
2
Heating
II
ALG
Two outputs are set for
the heater (J1-H1+H2
terminal block)
ALG
MB ADDR ZONE TEMPER
ATURE_TARGET
Table 20. Operating mode
Attempt to record a number beyond the allowable values leads to an error message. The variable is stored in
backed-up EEPROM memory.
MB_ADDR_FAN_SPEED – depending on operating condition (see
MB_ADDR_ZONE_MODE):
•
For „heating I” and „heating II” mode – variable that determines a fan rotation, that will be set in
case of setting the heater through two-state algorithm
•
For “air supply” mode – a variable that determines a current fan rotation.
Allowable values for the variable fall within a range from 0 to 3. The variable is stored in backed-up
EEPROM memory.
Fan rotation
Register from which the value will be
set at analogue output
0*
n/a
1
MB_ADDR_FAN_POWER_1_VOLT
2
MB_ADDR_FAN_POWER_2_V OLT
3
MB_ADDR_FAN_POWER_3_V OLT
* value record possible only in „supply” mode
Table 21: Dependency between a selected gear and occurrence of analogue output connected with a fan
Caution !!!
• Voltage at the analogue value calculated from the formula:
U
AO
=
(
MB_ADDR_FAN_POWER_x_VOLT
+
MB_ADDR_FAN_ADDITIONAL_V OLT)/ 100[V].
14