55
Follow the steps below to measure DC currents (A ):
- Switch on the DMM and select “A”, “mA”, or “µA” mode.
- The table shows the different modes and potential measurement ranges. Select the measurement range and cor-
responding measurement sockets.
Measurement mode Measurement range Measurement sockets
µA
0 - 6000 µA
COM + mAµA
mA
0 - 600 mA
COM + mAµA
A
0 - 10 A
COM + 10A
- Insert the red test lead into the mA µA or 10 A measurement socket. Insert the black test lead into the COM meas-
urement socket.
- Connect the two measuring probes (de-energised) in series to the object that you want to measure (e.g. battery or
circuit). The electrical circuit must be disconnected before you connect the probes.
- Reconnect the circuit. The measurement will appear on the display.
- After measuring, disconnect the circuit and remove the test leads from the measured object. Switch the DMM off.