c) Switching on/off

•  Briefly press the   button 

(4)

 to switch the measurement device on or off.

•  When you switch it on, all elements will appear on the LCD display 

(5)

 for a few seconds. 

The measured value is then displayed.

•  The measuring device switches off automatically after approximately 5 minutes of inactivity.

d) Calibration

This measuring device does not need to be calibrated. However, the measurement probe 

(7)

 

can be tested for normal functionality using a special ORP solution.

e) Making measurements

1.  Remove the probe cap 

(8)

.

2.  Press the   button 

(4)

 to switch on the measurement device.

3.  Dip the tip of the measuring probe 

(7)

 approx. 2 - 3 cm into the liquid that you want to 

measure. It may take a while for the measurement to stabilize. Read the measurement on 

the LCD display 

(5)

.

4.  If the measured value is outside of the measurement range, "----" will appear on the LCD 

display. 

5.  Rinse the measurement probe with distilled water after use.

f) Changing the measurement probe (7)

1.  Loosen the screw cap 

(3)

 in a clockwise direction.

2.  Remove the used/damaged probe 

(7)

 from the measuring device.

3.  Attach the new probe to the meter.
4.  Turn the screw cap anticlockwise until it is securely in place.

Care and cleaning

a) General cleaning

•  This product does not require any maintenance, except for changing the batteries and 

cleaning the measurement probes. Do not attempt to disassemble the product (except for 

the actions described in these instructions, such as changing the measurement probes).

•  Clean the product carefully using a soft, clean cloth. Do not press too hard on the display, 

as this may result in scratch marks or damage its function.

•  Only use distilled water (e.g. deionized water) to clean the measurement probes. Using 

other liquids may affect the measurements. Only use paper towels to dry the measurement 

probes.

•  Never use aggressive detergents, rubbing alcohol or other chemical solutions, as these 

may damage the casing or cause the product to malfunction.

b) Cleaning and polishing the sensor

If the measurements start to become inaccurate, follow the steps below to clean the sensor:
•  Remove the sensor cap 

(1)

.

•  Moisten the sand paper with some water.
•  Turn over the measuring device.
•  Using the pressure of your thumb, rub the platinum electrode with the wet sand paper about 

10 times in order to polish it.

•  Replace the sensor cap and rinse everything with clean, distilled water.

 

 

Disposal

a) Product

  Electronic devices are recyclable waste and must not be placed in household waste. 

At the end of its service life, dispose of the product according to the relevant statu-

tory regulations.

 

Remove any inserted batteries and dispose of them separately from the product.

b) Batteries

  You are required by law to return all used batteries (Battery Directive). They must not 

be placed in household waste.

 

Batteries that contain hazardous substances are labelled with this symbol to indicate 

that disposal in domestic waste is forbidden. The abbreviations for heavy metals in 

batteries are: Cd = Cadmium, Hg = Mercury, Pb = Lead (indicated on the battery, 

e.g. below the waste icon on the left).

Used batteries can be returned free of charge to local collection points, our stores or battery 

retailers.
You thus fulfil your statutory obligations and contribute to environmental protection.

Technical data

Power supply ................................ 2x 3 V/DC lithium batteries
Power consumption ...................... 5 mA
Standby ........................................ 5 µA
Battery life ..................................... Approx. 40 hours (continuous operation)
LCD display .................................. 21 x 18 mm
Measurement range ..................... ± 1999 mV
Accuracy ....................................... ±2 % FS
Resolution ..................................... 1 mV
Protection rating ........................... IP57
Operating/storage conditions ........  0 to +50 ºC, <85 % 

relative humidity (non-condensing)

Dimensions (Ø x H) ...................... 37 x 168 mm (measuring device)
Weight ..........................................  332 g (complete product) 

85 g (measuring device with batteries)

This is a publication by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240 Hirschau (www.conrad.com). 
All rights including translation reserved. Reproduction by any method (e.g. photocopying, microfilming or capturing in elec-

tronic data processing systems) requires prior written approval from the publisher. Reprinting, also in part, is prohibited. 

This publication reflects the technical status at the time of printing.
Copyright 2019 by Conrad Electronic SE.  

*1763393_v1_0819_02_DS_m_4L_(1)

Содержание ORP-01

Страница 1: ...Redox Messstick ORP 01 Best Nr 1763393 Bestimmungsgem e Verwendung Das Produkt ist f r die Messung des Redox Potentials ORP unter Verwendung von geeig neten Elektroden in Wasser und w ssrigen Medien a...

Страница 2: ...n Fall aggressive Reinigungsmittel Reinigungsalkohol oder an dere chemische L sungen da dadurch das Geh use angegriffen oder gar die Funktion beeintr chtigt werden kann b Sensor s ubern und polieren R...

Страница 3: ...measurement probe is not covered by the warranty guarantee Operating instructions Redox measurement stick ORP 01 Item no 1763393 Intended use This product is designed to be used in combination with su...

Страница 4: ...If the measurements start to become inaccurate follow the steps below to clean the sensor Remove the sensor cap 1 Moisten the sand paper with some water Turn over the measuring device Using the press...

Страница 5: ...r mesurer le potentiel Redox ORP dans l eau et les milieux aqueux au moyen d lectrodes ad quates Les exemples d applications de ce produit incluent eau potable eaux us es eaux superficielles piscines...

Страница 6: ...queraient d tre fauss s Pour s cher la pointe de mesure tapotez la d licatement avec une serviette en papier et rien d autre N utilisez en aucun cas des produits de nettoyage abrasifs de l alcool ou d...

Страница 7: ...garantie Gebruiksaanwijzing Redox meetstick ORP 01 Bestelnr 1763393 Doelmatig gebruik Het product is voor de meting van de redoxpotentiaal ORP met gebruik van geschikte elek troden in water en watera...

Страница 8: ...op de hieronder beschreven manier Ga hiervoor als volgt te werk Verwijder de sensorkap 1 Bevochtig het schuurpapier met wat water Draai het meetapparaat om Wrijf met het vochtige schuurpapier over de...

Отзывы: