99
Probe Specifications
DSO-4022/4042 Probe
Applicable model &
probe
DSO-4022, DSO-4042
GTP-060A-4
Position x 10
Attenuation Ratio 10:1
Bandwidth
DC
~
60MHz
Input Resistance 10M when used with 1M input
Input Capacitance 30pF approx.
Maximum Input
Voltage
600Vpeak CATII
Derating with frequency
Position x 1
Attenuation Ratio 1:1
Bandwidth
DC
~
6MHz
Input Resistance 1M when used with 1M input
Input Capacitance 200pF approx.
Maximum Input
Voltage
200Vpeak CATII
Derating with frequency
Operating Cond.
Temperature
–10°C ~ 55°C
Relative Humidity 85% @35°C
Safety Standard
EN 61010-031 CAT II
DSO-4062/4102 Probe
Applicable model &
probe
DSO-4062
GTP-060A-2
DSO-4102
GTP-100A-2
Position x 10
Attenuation Ratio 10:1
Bandwidth
DC
~
60MHz
DC
~
100MHz
Input Resistance 10M when used with 1M input
Input Capacitance 23pF
approx.
17pF
approx.
Maximum Input
Voltage
500V CAT I, 300V CAT II (DC+Peak AC)
Derating with frequency
Position x 1
Attenuation Ratio 1:1
Bandwidth
DC
~
6MHz
Input Resistance 1M when used with 1M input
Input Capacitance 180pF
approx.
47pF
approx.
Maximum Input
Voltage
300V CAT I, 150V CAT II (DC+Peak AC)
Derating with frequency
Operating Cond.
Temperature
–10°C ~ 55°C
Relative Humidity 85% @35°C
Safety Standard
EN 61010-031 CAT II
DSO-4000 User Manual
100
Declaration of Conformity
The below mentioned products
Type of Product: Digital Storage Oscilloscope
Model Number: DSO-4022, DSO-4042, DSO-4062, DSO-4102
are herewith confirmed to comply with the requirements set out in
the Council Directive on the Approximation of the Law of Member
States relating to Electromagnetic Compatibility (2004/108/EC) and
Low Voltage Directive (2006/95/EC).
For the evaluation regarding the Electromagnetic Compatibility and
Low Voltage Directive, the following standards were applied:
EMC
EN 61326-1: Electrical equipment for measurement, control and
laboratory use –– EMC requirements (2006)
Conducted Emission
Radiated Emission
EN 55011: Class A 1998 + A1:1999
+ A2:2002
Electrical Fast Transients
EN 61000-4-4: 2004
Current Harmonics
EN 61000-3-2: 2000 + A2:2005
Surge Immunity
EN 61000-4-5: 1995 + A1:2001
Voltage Fluctuations
EN 61000-3-3: 1995 + A1:2001 +
A2:2005
Conducted Susceptibility
EN 61000-4-6: 1996 + A1:2001
Electrostatic Discharge
EN 61000-4-2: 1995 + A1:1998 +
A2:2001
Power Frequency Magnetic Field
EN 61000-4-8: 1993 + A1:2001
Radiated Immunity
EN 61000-4-3: 2002 + A1:2002
Voltage Dip/ Interruption
EN 61000-4-11: 2004
Safety
Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
IEC/EN 61010-1: 2001