91
7. The Channel1 calibration
will complete in less than 5
minutes.
Ch1 calibration 1/3
8. When finished, connect the
calibration signal to the
Channel 2 input and repeat
the procedure.
Done!!
9. The calibration is completed and the display
goes back to the previous state.
Probe Compensation
Procedure
1. Connect the probe between Channel1 input and
the probe compensation output (2Vp-p, 1kHz
square wave) on the front panel. Set the probe
attenuation to x10.
CH1
x1
x10
2. Press the Utility key.
Utility
3. Press
ProbeComp
.
4. Press
Wavetype
repeatedly to
select the standard square
wave.
DSO-4000 User Manual
92
5. Press the Autoset key. The
compensation signal will
appear in the display.
Autoset
6. Press the Display key, then
Type
to select the vector
waveform.
7. Turn the adjustment point on the probe until
the signal edge becomes sharp.
Over
Compensation
Normal
Under
Compensation