49
14. Technical data
Power supply ....................................................9 V block battery
Response time ..................................................150 ms
Emission level ...................................................0.1 – 1.00 (adjustable)
IR measurement range .....................................-50 to +1200 °C (-58 to +2192 °F)
Contact measurement range ............................-50 to +1370 °C (-58 to +2498 °F)
Resolution .........................................................0.1 °C/°F
IR optics ............................................................50:1
Laser .................................................................output <1 mW, laser class 2, wavelength 630 - 670 nm
Operating conditions .........................................
0 to +50 °C, 10 – 90 % RH
Storage conditions ............................................
-10 to +60 °C, <80 % RH
System requirements ........................................Windows
®
2000 and higher
Weight ..............................................................330 g
Dimensions (W x H x D) ...................................204 x 52 x 155 mm
Measuring tolerances
The accuracy is valid for one year at a temperature of +23 to +25 °C (+73 to +77 °F), and at a relative humidity of
less than 75%, non-condensing.
Infrared temperature measurement range
Measurement range °C
Accuracy
Reproducibility
-50 to +20 °C
± 3 °C
± 1.5 °C
+20 to +500 °C
±1 % ±1 °C
±0.5 % or ±0.5 °C
+500 to +1000 °C
± 1.5 %
+1000 to +1200 °C
± 2.0 %
± 1.0 %
Measurement range °F
Accuracy
Reproducibility
-58 to +68 °F
± 5.4 °F
± 2.7 °F
+68 to +932 °F
±1 % ±1.8 °F
±0.5 % or ±0.9 °F
+932 to +1832 °F
± 1.5 %
+1832 to +2192 °F
± 2.0 %
± 1.0 %