65 7120
Elektr. Digital
Oberflächen Rauheitsmessgerät
Bedienungsanleitung
Ausstattung:
1. Ein/Aus-Schalter
2. LC Display
3. Funktions-Taste (mm/inch)
4. Funktions-Taste/Kalibrierung (Ra/z)
5. Schutzabdeckung
6. Messkopf
7. Batterie-Ladebuchse
8. Start-Taste
9. Reset-Taste
Spezifikation:
- Messungsparameter (µm):
Ra, Rz, Rq, Rt
- Prüfzuglänge (mm):
6
- Musterlänge (mm):
0,25 / 0,80 / 2,50
- Bewertungslänge (mm):
1,25 / 4.0
- Messbereich (µm):
Ra: 0,05 ~ 10,0
Rz: 0,1 ~ 50
- Anzeigefehler:
± 15%
- Abweichung der Anzeige:
<12 %
- Radius der Tastenspitze und Sensorwinkel:
10 µm ± 1 µm
Winkel: 90 +5° / -10°
- Messkraft der Tastspitzen:
≤ 0,
016N
Geschwindigkeit der Messkraft:
≤
800 N/m
- Druck des Sensorkopfes:
≤
0,5 N
- Batterie:
3,7 V, Lithium Batterie
- Abmessung:
106mm x 70mm x 24mm
- Gewicht:
200g
- Arbeitstemperatur:
-20°C ~ 40°C
- Luftfeuchtigkeit:
<90% RH
Lieferumfang:
-
Messgerät
-
USB-Ladekabel
-
Stecker
-
Kalibrier-Platte (Rauheitsnormal)
-
Bedienungsanleitung
-
Kunststoff-Tragekoffer
65 7120
Electronic Digital Surface
Roughness Measuring Device
Operation manual
Charge battery
Connect the charge cable to the roughness measuring device. Insert the plug in the respective socket (7) of the
roughness measuring device. The charger then has to be connected with the mains or a computer. Charging
duration is about 3 hours. Measurements can still be carried out during the charging procedure.
Device care
-
Protect the device against impacts and vibrations, excessive dust, moisture, oil
stains and strong magnetic fields, etc.
-
If you are not using the device, please switch it off in good time in order to keep the energy consumption as
low as
possible and charge within due time to avoid damage.
-
The sensor / measuring head is the precision unit of the testing device and should be treated with special
caution. After each test, cover the measuring head carefully with the protected area in order to avoid undesired
affects.
-
The enclosed standard plate for calibration should be carefully protected in order to avoid scratches that may
influence
the measuring accuracy.
-
In event of malfunctions, the user should under no circumstances open the device to attempt to make
repairs on it.
Send the device back to the manufacturer along with the guarantee card and the relevant probe as well as
a description so that it can be checked and repaired. To do so, please maintain contact with our customer
representatives or the sales department of our company.
Troubleshooting:
Error
Possible cause
Troubleshooting
No display
Battery is flat
Charge the battery
Battery symbol
appears in the
display
Battery is flat
Charge the battery
Please note:
In order to exclude incorrect measurements, please perform the basic setting carefully.
Clean the measuring instrument on a regular basis using a cotton cloth, do not use benzine or
solvent. Keep the measuring instrument dry as humidity may cause damage to it.
Electric etching or engraving may destroy the electronics, this shall void the guarantee!
Remove the battery if the device is not used for longer periods. Do not discard of used batteries in
domestic waste! Dispose of used batteries at designated collecting points or as special refuse.