Page 8
3.0 Factory
Setting
a)
Press [ON/OFF] button and the scale will perform a countdown.
Once the display shows ‘------’,
press the [ZERO] button and [UNIT] button at the same time.
b)
Scale display “XXX CAP”.
Press [UNIT] button to change the capacity range.
Press [ZERO] to confirm the setting.
c)
Scale display ‘X.XXX DIV’.
Press [UNIT] to change the division.
Press [ZERO] to confirm the setting.
d)
Scale display ‘X SEG’.
Press [UNIT] to change calibration point.
Press [ZERO] to confirm the setting.
e)
Scale display ‘XXXXXX S-0’.
Press [ZERO] when the stability sign appear.
f)
Press [ZERO] to go ‘XXXXXX S-1’.
Place test weigh (1 part of total capacity) on weighing pan.
When the scale display shows the stability sign, press [ZERO].
E.g. 100g when the scale shows 1.XXXXX
E.g. 200g when the scale shows 2.XXXXX
E.g. 300g when the scale shows 3.XXXXX