background image

 

 

Use  only 

VivaChek™  Ino/Fad  test  strip  with  your  VivaChek™  Fad  Sync 

meter.   

 

Use  only 

VivaChek™  Ino/Fad  control  solution  with  your  VivaChek™  Fad 

Sync meter and 

VivaChek™ Ino/Fad strip.   

 

Please  contact  your  physician  or  diabetes  healthcare  professional  if  you 
determine to make a change on your current medical therapy or diet activity 
based on test results.   

 

If the  system  is used  in  a  manner  not  specified  by  the  manufacturer, 
the protection provided by the system can be impaired. 
 

 

Potential Biohazard

 

All  parts  of  the  kit  are  considered  biohazardous  and  can  potentially  transmit 
infectious diseases, even after you have performed cleaning and disinfection. 

 

Note:

 

1.  The  meter  and  lancing  device  are  for  single  patient  use.  Do  not  share  them 

with anyone including other family members! Do not use on multiple patients! 

2.  All parts of the kit are considered biohazardous. They can potentially transmit 

infectious  diseases  from  blood  borne  pathogens,  even  after  you  have 
performed  cleaning  and  disinfection.  Please  follow  proper  precautions  when 
handling your meter and lancing device. 

3.  For more information, please refer to the FDA Public Health Notification: 

“Use 

of Fingerstick Devices on More than One Person Poses Risk for Transmitting 
Bloodborne 

Pathogens: 

Initial 

Communication” 

(2010) 

at 

http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm. 
You may also refer to the CDC Clinical Reminder: 

“Use of Fingerstick Devices 

on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” 
(2010) at http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html. 

 
Limitations 

 

For single-patient use only. 

 

When  using  with 

VivaChek™  Fad  test  strips,  very  high  (above  70%) 

hematocrit levels can cause false results. 

 

When  using  with 

VivaChek™  Ino  test  strips,  very  high  (above  70%)  and 

very low (below 20%) hematocrit levels can cause false results. Talk to your 
health care professional to find out your hematocrit level. 

 

If you are taking vitamin C (ascorbic acid in your blood > 3 mg/dL) then your 
glucose results using this meter may not be reliable. 

 

Patients undergoing oxygen therapy may cause false result. 

 

Not for use on patients with critical illness.   

 

Not for use on patients in shock, or with severe dehydration or from patients 

Содержание Fad Sync

Страница 1: ...Chek Fad Blood Glucose Test Strips VGS02 The VivaChek Fad Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples dr...

Страница 2: ...results wirelessly to the app 9 Using Your Meter without App 9 3 Taking a Test 10 Preparing the Test Strip 10 Getting a Blood Drop and Testing 12 Discard the Used Test Strip 13 Testing with Control S...

Страница 3: ...1 CHARPTER 1 UNDERSTANDING YOUR TESTING TOOLS Your Meter System Overview The VivaChek Fad Sync Blood Glucose Meter and VivaChek Ino Fad Strip...

Страница 4: ...ly like the picture below If not contact VivaChek Customer Support at 800 951 8430 7 days a week 9 a m 5 p m Eastern Time for help Please contact your health care professional if you need help outside...

Страница 5: ...meter by running a quality control test Refer to Quality Control Test for instructions Do not take the meter apart This will void the warranty Refer to the Caring for Your Meter section for details on...

Страница 6: ...roper precautions when handling your meter and lancing device 3 For more information please refer to the FDA Public Health Notification Use of Fingerstick Devices on More than One Person Poses Risk fo...

Страница 7: ...antenna or transmitter This meter has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable...

Страница 8: ...mode Press to adjust it then press and hold to save your choice Start setting the year month and date 2 Set the Date The year will now flash Press to adjust it then press and hold until the meter bee...

Страница 9: ...you first time use your meter system for testing please adjust the meter settings to set the date and time ensuring that results stored in the memory are shown with the correct date and time 3 Set the...

Страница 10: ...the minute press and hold to set Set the Audio Feature After you set the time press to select On or OFF Press and hold to set Now you have completed your meter set up A symbol of a test strip appears...

Страница 11: ...rom tests taken before your initial Sync will never be sent to the app To pair the Smartphone with your meter turn your meter on and follow these steps 1 When you turn on your meter the symbol will ap...

Страница 12: ...on after a beep 4 A symbol with a test strip with a flashing blood will appear letting you know the meter is ready to test Note Check the expiration and discard dates on the test strip vial All expira...

Страница 13: ...ncing device uses ONLY VivaChek sterile lancet 2 Hold the lancet firmly in the lancing device and twist the safety tab of the lancet until it loosens then pull the safety tab off the lancet Save the s...

Страница 14: ...indicate the lancing device is now loaded and ready for obtaining a drop of blood 2 Press the lancing device against the side of the finger to be lanced with the cover resting on the finger Push the...

Страница 15: ...mple has been fully filled the check window on the strip tip Hold the tip of the test strip in the blood drop until the meter beeps Note If the blood sample does not fill the check window do not add a...

Страница 16: ...than the fingertip Because of this you may find that obtaining blood from these sites is less painful than from the fingertip The procedure for forearm and palm sampling is different You need the lanc...

Страница 17: ...d dry away from bone and free of visible veins and hair To bring fresh blood to the surface of the puncture site massage the puncture site vigorously for a few seconds until you feel it getting warm 5...

Страница 18: ...ANT There are important differences among forearm palm and fingertip samples that you should know Important Information about forearm and palm glucose testing You should consult your doctor before cho...

Страница 19: ...1 8430 7 days a week 9 a m 5 p m Eastern Time Please contact your health care professional if you need help outside of these hours Testing with Control Solution Why Perform Control Tests Performing a...

Страница 20: ...n image of a test strip with a flashing blood drop will appear letting you know the meter is ready to test 4 Shake the control solution bottle thoroughly Squeeze the control solution bottle gently and...

Страница 21: ...on the test strip vial label Notes If your control test result is out of range Check the expiration dates and discard dates of the test strip and control solution Make sure that the test strip vial an...

Страница 22: ...just remove the test strip Dispose of the used test strips as medical waste The result will be automatically marked and stored in the meter memory Control results will be not included in your blood gl...

Страница 23: ...hot or has an unusual odour Contact your local distributor Do Not plug the AC adapter into a wall socket and leave it unattended Verify that the wall socket voltage matches the AC adapter voltage Do...

Страница 24: ...be performed prior to disinfection but cleaning does not kill germs After use and exposure to blood all parts of this kit can potentially transmit infectious diseases Disinfecting reduces the risk of...

Страница 25: ...ver the 5 year life of the meter Notes Do not use alcohol or any other solvent Do not allow liquid dirt dust blood or control solution to enter the test strip port or the data port Do not squeeze the...

Страница 26: ...test strip and repeat the test with a new test strip Make sure that you are using a VivaChek Ino Fad test strip from VivaChek Biotech Hangzhou Co Ltd Incorrect sample Discard the test strip and repeat...

Страница 27: ...is above 600 mg dL Wash and dry your hands well and the test site Repeat the test using a new test strip If your result still flashes HI contact your health care professional as soon as possible Cauti...

Страница 28: ...symptoms test your blood glucose If your test result is under 50 mg dL or above 250 mg dL contact your health care professional immediately Low blood glucose Hypoglycemia shakiness sweating fast hear...

Страница 29: ...per deciliter mg dL Memory Up to 500 records with date and time Automatic Shutoff 2 minutes after last action Dimensions 82 mm x 54 mm x 23 mm Display Size 32mm x 32 mm Weight Approximately 53g Opera...

Страница 30: ...x Accurate Results Meter result is 15 of laboratory result More Accurate Results Meter result is 10 of laboratory result Most Accurate Results Meter result is 5 of laboratory result Warranty Please co...

Отзывы: