background image

12 

 

 

Adjustment:   

1                for delicate skin   
2 and 3    for normal skin   
4 and 5    for calloused or thick skin 

Note:   
Greater pressure of the lancing device against the finger will also increase the 
puncture depth 

 

Getting a Blood Drop and Testing 

1.  Pull the cocking barrel back to set  the lancing  device. You may hear  a click; 

while the release button changes to orange color to indicate the lancing device 
is now loaded and ready for obtaining a drop of blood. 

 

2.  Press  the  lancing  device  against  the  side  of  the  finger  to  be  lanced  with  the 

cover resting on the finger. Push the release button to prick your fingertip. You 
should hear a click as the lancing device activates. 

 

3.  Gently massage from the base of the finger to the tip of the finger to obtain the 

required  blood  volume.  Avoid  smearing  the  drop  of  blood.  For  the  greatest 
reduction in pain, lance on the sides of the fingertips. Test immediately after a 
good blood drop has formed.   

Содержание Fad Sync

Страница 1: ...Chek Fad Blood Glucose Test Strips VGS02 The VivaChek Fad Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples dr...

Страница 2: ...results wirelessly to the app 9 Using Your Meter without App 9 3 Taking a Test 10 Preparing the Test Strip 10 Getting a Blood Drop and Testing 12 Discard the Used Test Strip 13 Testing with Control S...

Страница 3: ...1 CHARPTER 1 UNDERSTANDING YOUR TESTING TOOLS Your Meter System Overview The VivaChek Fad Sync Blood Glucose Meter and VivaChek Ino Fad Strip...

Страница 4: ...ly like the picture below If not contact VivaChek Customer Support at 800 951 8430 7 days a week 9 a m 5 p m Eastern Time for help Please contact your health care professional if you need help outside...

Страница 5: ...meter by running a quality control test Refer to Quality Control Test for instructions Do not take the meter apart This will void the warranty Refer to the Caring for Your Meter section for details on...

Страница 6: ...roper precautions when handling your meter and lancing device 3 For more information please refer to the FDA Public Health Notification Use of Fingerstick Devices on More than One Person Poses Risk fo...

Страница 7: ...antenna or transmitter This meter has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable...

Страница 8: ...mode Press to adjust it then press and hold to save your choice Start setting the year month and date 2 Set the Date The year will now flash Press to adjust it then press and hold until the meter bee...

Страница 9: ...you first time use your meter system for testing please adjust the meter settings to set the date and time ensuring that results stored in the memory are shown with the correct date and time 3 Set the...

Страница 10: ...the minute press and hold to set Set the Audio Feature After you set the time press to select On or OFF Press and hold to set Now you have completed your meter set up A symbol of a test strip appears...

Страница 11: ...rom tests taken before your initial Sync will never be sent to the app To pair the Smartphone with your meter turn your meter on and follow these steps 1 When you turn on your meter the symbol will ap...

Страница 12: ...on after a beep 4 A symbol with a test strip with a flashing blood will appear letting you know the meter is ready to test Note Check the expiration and discard dates on the test strip vial All expira...

Страница 13: ...ncing device uses ONLY VivaChek sterile lancet 2 Hold the lancet firmly in the lancing device and twist the safety tab of the lancet until it loosens then pull the safety tab off the lancet Save the s...

Страница 14: ...indicate the lancing device is now loaded and ready for obtaining a drop of blood 2 Press the lancing device against the side of the finger to be lanced with the cover resting on the finger Push the...

Страница 15: ...mple has been fully filled the check window on the strip tip Hold the tip of the test strip in the blood drop until the meter beeps Note If the blood sample does not fill the check window do not add a...

Страница 16: ...than the fingertip Because of this you may find that obtaining blood from these sites is less painful than from the fingertip The procedure for forearm and palm sampling is different You need the lanc...

Страница 17: ...d dry away from bone and free of visible veins and hair To bring fresh blood to the surface of the puncture site massage the puncture site vigorously for a few seconds until you feel it getting warm 5...

Страница 18: ...ANT There are important differences among forearm palm and fingertip samples that you should know Important Information about forearm and palm glucose testing You should consult your doctor before cho...

Страница 19: ...1 8430 7 days a week 9 a m 5 p m Eastern Time Please contact your health care professional if you need help outside of these hours Testing with Control Solution Why Perform Control Tests Performing a...

Страница 20: ...n image of a test strip with a flashing blood drop will appear letting you know the meter is ready to test 4 Shake the control solution bottle thoroughly Squeeze the control solution bottle gently and...

Страница 21: ...on the test strip vial label Notes If your control test result is out of range Check the expiration dates and discard dates of the test strip and control solution Make sure that the test strip vial an...

Страница 22: ...just remove the test strip Dispose of the used test strips as medical waste The result will be automatically marked and stored in the meter memory Control results will be not included in your blood gl...

Страница 23: ...hot or has an unusual odour Contact your local distributor Do Not plug the AC adapter into a wall socket and leave it unattended Verify that the wall socket voltage matches the AC adapter voltage Do...

Страница 24: ...be performed prior to disinfection but cleaning does not kill germs After use and exposure to blood all parts of this kit can potentially transmit infectious diseases Disinfecting reduces the risk of...

Страница 25: ...ver the 5 year life of the meter Notes Do not use alcohol or any other solvent Do not allow liquid dirt dust blood or control solution to enter the test strip port or the data port Do not squeeze the...

Страница 26: ...test strip and repeat the test with a new test strip Make sure that you are using a VivaChek Ino Fad test strip from VivaChek Biotech Hangzhou Co Ltd Incorrect sample Discard the test strip and repeat...

Страница 27: ...is above 600 mg dL Wash and dry your hands well and the test site Repeat the test using a new test strip If your result still flashes HI contact your health care professional as soon as possible Cauti...

Страница 28: ...symptoms test your blood glucose If your test result is under 50 mg dL or above 250 mg dL contact your health care professional immediately Low blood glucose Hypoglycemia shakiness sweating fast hear...

Страница 29: ...per deciliter mg dL Memory Up to 500 records with date and time Automatic Shutoff 2 minutes after last action Dimensions 82 mm x 54 mm x 23 mm Display Size 32mm x 32 mm Weight Approximately 53g Opera...

Страница 30: ...x Accurate Results Meter result is 15 of laboratory result More Accurate Results Meter result is 10 of laboratory result Most Accurate Results Meter result is 5 of laboratory result Warranty Please co...

Отзывы: