23
Step 2: Wipe the entire surface including front side, back side, right side and left
side of the meter, by a back and forth movement.
Step 3: Keep the meter surface wet for at least one minute.
Step 4: Wait for the surface of meter to be dry.
Clorox
TM
Healthcare Bleach Germicidal Wipes containing Sodium hypochlorite
0.55%, which has been proven to be safe to use with the VivaChek™ Fad Sync
system. Clorox
TM
Healthcare Germicidal Bleach Wipes are available by contacting
Krasity Medical Supply at 800-537-1394 directly or visiting and purchasing at
http://www.walmart.com and http://www.staples.com/.
The meter should be disinfected a minimum of once per week. The meter
disinfection process has been validated for 608 disinfection cycles, which is
equivalent cleaning and disinfecting your meter every 3 days for 5 years. This is to
ensure that your meter will operate properly over the 5-year life of the meter.
Notes:
•
Do not use alcohol or any other solvent.
•
Do not allow liquid, dirt, dust, blood, or control solution to enter the test strip port
or the data port.
•
Do not squeeze the wipe or gauze into test strip port.
•
Do not spray cleaning solution on the meter.
•
Do not immerse the meter in any liquid.
Notes:
Although it has not been observed, some alterations may appear on your meter
due to the cleaning and disinfection procedure. Such as: cloudy display window,
plastic housing cracking, meter buttons do not function, partial display on full
screen, unable t
o execute the meter’s initial set up, etc. If you notice any of these
external changes to your meter or any changes to the performance of your meter
stop using the meter and please contact Customer Support for help.
If you have questions about cleaning or disinfection, or if you see evidence of
physical damage, contact VivaChek Customer Support at 800-951-8430 (available
7 days a week, 8 a.m.- 5 p.m. Eastern Time).
Please contact your health care
professional if you need help outside of these hours.