12. 

Technical Specification

Product

Respiratory Monitor COPD-6

Model

4000

Dimensions

109mm (length) x 63mm (width) 

x 42mm (height)

Weight

63g (not including batteries)

Flow Detection Principal

Stator/rotor

Accuracy:

Better than ± 3%

Flow Impedance:

Better than 0.15kPa/L/s at 

14L/s

Back pressure

Less than 0.15kPa/L/second 

@ 14L/s

Measurement Range:

0 – 9.99 L BTPS

Power Supply:

3V (2 x 1.5V AAA batteries)

Battery Life

3 months of use, 3 tests per day
(Batteries near the end of their 

shelf life will have reduced 

capacity.)

! Bad Test Criteria:

Slow start of test (Vext>5%) or 

a cough detected in the first 

second

FEV substituted for FEV6

When FET < 3s and abrupt end 

of test

Product Life

6 Years for the main device. The 

Eco BVF, SafeTway & nose clips 

are 7 years.

Operating temperature range

17 – 37°C

Operating humidity range

30%–75%

Ambient pressure range

850hPa–1060hPa

Performance and Safety 

Standards:

ATS/ERS 2019, ISO 23747:2015, 

ISO 26782:2009

Safety standards

EN 60601-1, EN 60601-1-11

Содержание copd-6 4000

Страница 1: ...copd 6 Instructions for Use MODEL 4000 Copyright Vitalograph 2020 Current Edition Issue 1 28 Jan 2020 Cat No 09780...

Страница 2: ...t vitalograph de Vitalograph Inc 13310 West 99th Street Lenexa Kansas 66215 USA Toll Free 800 255 6626 Tel 913 730 3200 Fax 913 730 3232 E mail contact vitalograph com www vitalograph com Technical Su...

Страница 3: ...Vitalograph COPD 6 Early detection of COPD Displays FEV1 FEV6 ratio and predicted obstructive index COPD classification stage I IV and lung age Identifies risk of COPD at pre symptomatic stage Screens...

Страница 4: ...lograph intends that a new Bacterial Viral Filter Eco BVF be used for every subject to prevent cross contamination Using a new Eco BVF provides a significant level of protection of the subject the dev...

Страница 5: ...h the correct height Height increases decreases in values of 1 If the button is kept depressed the values scroll faster Press to save the height Note if height values are set below 100 the device assu...

Страница 6: ...a Breathe in as deeply as possible b Hold the breath c Place the mouthpiece into the mouth biting the mouthpiece lightly and sealing the lips firmly around it d Blow out as HARD and FAST as possible...

Страница 7: ...ssing again will toggle between best and last blow 13 Pressing the button shows the FEV6 and FEV6 Predicted results 14 Pressing the button again shows the FEV1 FEV6 and FEV1 FEV6 Predicted results 15...

Страница 8: ...lour systems for each zone type are pre set as follows FEV1 Pred Obstructive Index COPD Classification FEV1 FEV6 Ratio and FEV Pred 80 0 Green Not COPD FEV1 FEV6 0 7 Top Boundary Stage I FEV1 FEV6 0 7...

Страница 9: ...bottom boundary by pressing the or button and releasing when the value is reached The values increase decrease in values of 1 If the button is kept depressed the values scroll faster 8 Press to save...

Страница 10: ...used Weekly cleaning of the mouthpiece outside surfaces and flowhead of the device is recommended A cloth impregnated with 70 isopropyl alcohol may be used The plastic mouthpiece may be washed in warm...

Страница 11: ...ting up using or maintaining the device or to report unexpected operations or changes in performance contact Vitalograph using the contact information at the start of this manual Also contact the heal...

Страница 12: ...product end of life Do not dispose of these products as unsorted municipal waste The pouch can be disposed of in unsorted municipal waste Used Eco BVFs and SafeTways constitute minimally soiled waste...

Страница 13: ...evice Order Number Use by Date Date format yyyy mm dd Keep Dry Do not re use Non sterile Recycle QR code matrix bar code All information in the bar code is included in the text under it Other Labels B...

Страница 14: ...ce displays some key parameters for clinical interpretation FEV1 and FEV1 Predicted FEV6 and FEV6 Predicted FEV1 FEV6 and FEV1 FEV6 Predicted If an abnormal result is indicated arrows on the display s...

Страница 15: ...batteries Battery Life 3 months of use 3 tests per day Batteries near the end of their shelf life will have reduced capacity Bad Test Criteria Slow start of test Vext 5 or a cough detected in the firs...

Страница 16: ...F be used for every subject to prevent cross contamination Using a new Eco BVF provides a significant level of protection of the subject the device and the user against the risk of cross contamination...

Страница 17: ...enna cables and external antennas which emit electromagnetic fields should be used no closer than 30 cm 12 inches to any part of the device including cables specified by Vitalograph Otherwise degradat...

Страница 18: ...The customer or the user of the COPD 6 should assure that it is not used in such an environment The Model 4000 COPD 6 has been tested in accordance with EN60601 1 2006 A1 2013 Medical electrical equi...

Страница 19: ...tablishments EN 60601 1 2 Immunity tests Immunity test Test level Compliance level Reached Electrostatic discharge ESD EN 61000 4 2 Contact 8 kV Air 2 kV 4 kV 8kV 15 kV Contact 8 kV Air 2 kV 4 kV 8kV...

Страница 20: ...nterference is suspected or possible switching off the offending device is the normal solution as is required in aircraft and medical facilities Medical electrical equipment needs special precautions...

Страница 21: ...mended This device is classified as IIa per Annex IX of the MDD also meets the provisions of the Essential Requirements Annex I via compliance with Annex II of the Medical Devices Directive as per Art...

Страница 22: ...akes to rectify at no expense to the customer any software failure notified within the period stated above provided that the failure can be recreated and the software has been installed and used in ac...

Страница 23: ...y consequential damages arising out of the use of or inability to use any Vitalograph equipment 8 This Guarantee is offered as an additional benefit to the Consumer s statutory rights and does not aff...

Страница 24: ......

Отзывы: