EN 60601-1-2 - Emissions tests
Emissions test
Compliance
Electromagnetic
environment - guidance
RF emissions
CISPR 11
Group 1
The Model 4000 COPD-6
uses RF energy only for its
internal function. Therefore,
its RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Model 4000 COPD-6
is suitable for use in all
establishments, including
domestic establishments.
EN 60601-1-2 Immunity tests
Immunity test
Test level
Compliance level
Reached
Electrostatic
discharge (ESD)
EN 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Radiated RF
EN 61000-4-3
3 V/m
80MHz to
2700MHz
3 V/m
80MHz to 2700 MHz
Proximity fields
from RF devices
EN 61000-4-3
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
Medical Devices may be affected by mobile RF communications
equipment including cellular telephones and other personal
or household devices not intended for medical facilities. It is