Additional error message from the VITA ZYRCOMAT 6000 MS
Display
Recognition
Cause/correction
Ventilator power board defective.
Current program is stopped, heating off.
• Inspect connections to the circuit
board.
• Inspect/replace ventilator.
Temperature sensor defective.
Running program is stopped, heating off.
• Inspect connections to the circuit
board.
• Inspect/replace ventilator.
Relay error.
Running program is stopped, heating off.
• Replace power board.