VI. MODALITIES AND START-UP OF THERMOFLASH® LX-260T EVOLUTION
1. Install the batteries.
2. When the batteries are used for the first time or when inserting new batteries, wait about 15
minutes for the device to reach room temperature.
3. Aim the device at the temporal lobe or a side/temple, as shown (SEE DIAGRAM
A
PAGE 2), at
a distance of 2 to 5 cm (0.79 - 1.97 in) press the measurement key, the temperature is instantly
displayed. The temperature can also be taken behind the ear lobe - on the jugular vein.
VII. MENU SETTING & FUNCTIONS
1. Choice of temperature display mode
Body mode: Measurement of the body temperature.
Measuring range: 32°C to 42.9°C (89.6°F to 109.2°F)
Surface mode: Measurement of the surface temp of an object or food.
Measuring range: 0°C to 93.3°C (32.0°F to 199.94 °F)
Room mode: Measure the ambient temperature of a room.
Measuring range: 0°C to 50°C (32°F to 122°F)
Press the measurement button to turn on the device.
Press MODE button to select the temperature measurement mode.
The screen displays: Body mode - for body temperature measurement. Press the MODE button
again to switch to SURFACE mode. The screen displays Surface.
Press the MODE button again to switch to ROOM mode. The screen displays Room.
Note: the
THERMOFLASH LX-260T EVOLUTION
is automatically set to BODY.
IMPORTANT: The area temperature differs from the internal body temperature.
Use BODY mode only to obtain the internal body temperature.
2. Choice of temperature unit
Press the °C/°F button to select the temperature unit. The screen, displays the symbol of the
selected unit (°C for degrees Celsius, °F for degrees Fahrenheit).
3. Data memory
Press MEM for two seconds to display the last measure. Press MODE and C°/F° to scroll
through the last recorded measurements. To exit the menu, press the measure key. The device
automatically switches off after 5 seconds of inactivity.
4. Choosing the language
Use the Language key to choose a language for the vocal function. The choice of language is
visible.
Languages available: French, English, Spanish, Italian, German, Dutch, Silent Function. By default
the
ThermoFlash® LX-260T EVOLUTION
is in French.
5. Alarm threshold
The alarm threshold is set to 38°C. This cannot be changed. If the temperature is over 38°C, a
beep sounds during the temperature measurement.
Содержание THERMOFLASH LX-260T EVOLUTION
Страница 2: ...0197 G01079 ...
Страница 6: ...FR 0197 G01079 MANUEL D UTILISATION ...
Страница 14: ...0197 G01079 EN USER MANUAL ...
Страница 22: ...0197 G01079 NL HANDLEIDING ...
Страница 30: ...0197 G01079 ES MANUAL DE UTILIZACIÓN ...
Страница 38: ...0197 G01079 DE GEBRAUCHSANWEISUNG ...
Страница 46: ...0197 G01079 IT MANUALE D USO ...
Страница 54: ...53 IFU_G01079_ML_ V02202006R02 Visiomed LX 260TE Manuale d uso 032016 E M C D ATA ...
Страница 61: ...IFU_G01079_ML_ V02202006R02 60 ...
Страница 62: ...61 IFU_G01079_ML_ V02202006R02 Visiomed LX 260TE Manuale d uso 032016 WA R R A N T Y ...