IFU_G01079_ML_ V02202006R02
16
- Do not drink hot or cold drinks, and do not do strenuous exercise during temperature taking.
- The protective glass of the lens is the most important and fragile part of the thermometer, treat
it carefully and do not touch it with your fingers.
Constraints
In order to obtain a stable and reliable result, before taking the temperature:
- Move hair aside.
- If necessary, wipe away any perspiration from the forehead.
- Avoid drafts/any significant air flow (nasal cannula, air conditioning ...).
- Observe a 3-minute interval between two measures.
The reliability of the measure is not assured in case of temperature measurement on another
anatomical region (e.g. arms, torso ...).
IV. OPERATING PRINCIPLE
Each solid, liquid or gaseous body, has the property of emitting energy by radiation with
an intensity that depends on its temperature. The
THERMOFLASH® LX-260T EVOLUTION
infrared thermometer measures a persons temperature by means of the energy they emit.
This measurement is made possible by the instrument's external temperature probe, which
permanently analyses and records the ambient temperature. In such a way that as soon as
the operator holds the thermometer close to the body and activates the radiation sensor, the
measurement is carried out instantaneously by detection of the infrared heat emitted by the
arterial blood flow. The result of this body temperature measurement is then obtained without
interference from the heat of the surrounding environment.
INFORMATIVE TABLES
Normal temperatures according to procedure
MEASUREMENT MODE
NORMAL T°
Rectal
36.6°C – 38°C /
97.88°F – 100.4°F
Oral
35.5°C – 37.5°C /
95.9°F – 99.5°F
Axillary
34.7°C – 37.3°C /
94.46°F – 99.14°F
Aural
35.8°C – 38°C /
96.44°F – 100.4°F
Temporal (Thermoflash®)
35.8°C – 37.8°C /
96.44°F – 100.04°F
Chamberlain, J.M., Terndrup, T.E., New Light on Ear Thermometer - Readings, Contemporary Pediatrics, March 1994.
Chamberlain, J.M., et al., Determination of Normal Ear - Temperature with an Infrared Emission Detection Thermometer
Содержание THERMOFLASH LX-260T EVOLUTION
Страница 2: ...0197 G01079 ...
Страница 6: ...FR 0197 G01079 MANUEL D UTILISATION ...
Страница 14: ...0197 G01079 EN USER MANUAL ...
Страница 22: ...0197 G01079 NL HANDLEIDING ...
Страница 30: ...0197 G01079 ES MANUAL DE UTILIZACIÓN ...
Страница 38: ...0197 G01079 DE GEBRAUCHSANWEISUNG ...
Страница 46: ...0197 G01079 IT MANUALE D USO ...
Страница 54: ...53 IFU_G01079_ML_ V02202006R02 Visiomed LX 260TE Manuale d uso 032016 E M C D ATA ...
Страница 61: ...IFU_G01079_ML_ V02202006R02 60 ...
Страница 62: ...61 IFU_G01079_ML_ V02202006R02 Visiomed LX 260TE Manuale d uso 032016 WA R R A N T Y ...