Troubleshooting
Checking Load Cell Connection
VT300 Technical Manual, Rev. A4
80
Doc # TM-VT300-EN
10.1.4
Alibi Memory and Accumulated Total Errors
Error code & desc.
Possible cause
Actions to be taken
E55:ALIBI FULL
Alibi memory is full.
Press
ENTER
to acknowledge. The unique
identification number will reset to
0000
.
E56:NO ALIBI
Tare is active, so it is not
possible to print.
Printout aborted. Cancel tare by pressing
ESC
, and retry.
E57:ALIBI CSUM
Alibi memory is corrupted.
No action. The alibi memory cannot be
cleared, but the next records will be
corrected.
E67:TM CHECKSUM
Corrupted accumulated total.
Clear accumulated total (
TFILE
menu).
E69:TM OVERFLOW
Accumulated total has
overflowed because it has not
been cleared for a long time.
Clear accumulated total (
TFILE
menu).
E77:WM CHECKSUM
Corrupted weighing memory.
Use menu path
VFILE
>
DELETE
to clear
weighing memory.
10.2 Checking Load Cell Connection
If there appears to be a problem with the load cell connection:
Check input and output resistance.
Check resistance between any terminal and shield.
Check load cell connection and cable.
10.3 Checking Power Supply
If the unit does not turn on:
Check 9-15VDC power supply.
Check the resettable fuse F4 on PCB 801.
10.4 Checking Digital Outputs
If the setpoints are not working properly:
Test the setpoints (see section
8.2.3).
Check 24VDC power supply.
Check the resettable fuse F3 on PCB 801.