General System Parameters
SYSTEM > SET > OPER
VT300 Technical Manual, Rev. A4
41
Doc
#
TM-VT300-EN
5.2.1
Dual-scale Connecting And Parameter Settings
To connect and set up VT300 for dual-scale operations:
1. Set parameter
2SCL ENABLE(00)
to
YES
,
DUAL RANGE(19)
to 00, and
CONV.RATE
(07)
to a value less than 70.
2. Ascertain that the first scale is calibrated and working properly.
3. Connect the second scale to
ST5
, 7 pins socket in the one side and in the other
side using the mounting posts/spacers provided to:
Stainless Steel Enclosure
– The main board.
Aluminum Enclosure
– To the indicator’s rear panel.
4. Ascertain that the Scale 2 values for parameters
DEC.POINT (02)
and
SCALE.DIV (03)
are the same as those for Scale 1.
5. After checking that the
S2
annunciator is on,(switching with the
SCALE
key)
calibrate the second scale.
6. The device is now ready to work in dual-scale mode.
7. Use the
SCALE
key for switching between Scale1(
S1
on the display) , Scale2(
S2
on the display) or sum of Scale1 and Scale2(
S0
on the display).
5.3 SYSTEM > SET > OPER
# Parameter
Description
Values
01 TOTALIZING
M
Enable totalizer
. YES enables the accumulated total
function for users. NO disables it.
YES, NO
02 WEIGHING
MEM
Store first weighing in memory.
If set to NO, first
weighing information is discarded and cannot be accessed
through the
FIRST
submenu in the
VFILE
menu.
YES, NO
03 REFR
CODE
Number of descriptions for weighbridge operations.
Specifies the number of descriptions (reference codes) the
user is prompted for during first and second weighings. By
default these are VEHICLE, CLIENT and PRODUCT.
If set to 0, weighbridge descriptions are disabled
altogether. In any event, only the first two descriptions are
stored in memory.
0-4
0=disable
descriptions
04 CODE
MEM
Remember last weighbridge description.
Specifies
whether or not the first and second weighing dialogs
should display, for each description such as VEHICLE and
CLIENT, the value entered by the user in the previous
weighing operation.
YES, NO