eco-CONTROL EC200-B
Operation and maintenance instructions
7
Troubleshooting / maintenance
Troubleshooting
Maintenance
The Dosing control can be regarded as maintenance-free. Do not use any aggressive
solvents or cleaners for cleaning, only a damp cloth. Unplug the unit from the mains
before doing any cleaning.
Error
Possible cause
Correction
Dosing system cannot be
operated, no display.
no dispenser connected
connect dispenser (ensure that
the mains plug has been
removed beforehand)
Power supply off
switch on
Dosing program cannot be
saved, only program 00 is
offered
No chipcard inserted, or not
formatted
Insert / format the chipcard
Error messages
Overcurrent monitoring
Dispenser components do not
run smoothly due to hardened
medium or as a result of dry
running.
Dismantle and clean the
dispenser, replace the stator if
necessary.
See the operating instructions for
the dispenser.
Level of medium critical
Not enough medium in the
supply tank
Fill up with the medium. If the
error message remains, check
the sensor and sensor input, if
applicable, brief "emergency
operation" without level
monitoring.
Check the compressed air Compressed air too little, or
switched off
Check / switch on the
compressed air
Tolerance range set too small
Set the tolerance range to be
larger, see (5.10.6)
Primary pressure monitoring
has been switched on and the
dummy plug has not been
inserted into the sensor input.
Insert the dummy plug or turn off
the primary pressure monitoring,
see (5.10.10)