flowscreen
Operating Instructions
18 / 24
Version 1.0
Copyright ©
3.6
Performing measurements
Measurement operation requires the settings from the following sections:
- 3.4, Starting up
- 3.5, Base settings
The flowscreen analysis system is then ready for operation (to switch on, see 3.3). If
envelope curves have been saved (see 3.5.5), these must be reset every time after
switching on.
Notice: The following description applies to both sensor inputs, which can each be
configured differently from each other.
Measurement display
The pressure / time gradient of each measurement (pressure curve) is shown on the
display.
Error display
If one or more of the set values / tolerances is exceeded in a measurement, this is shown
optically (red)
Error counter
The error counter increases by a value of 1 each time an error occurs. The error LED
illuminates as soon as the set threshold value is reached.
Error-free dosings increase the respective counter in the same way.
Error reset
<
Reset Counter
> (diagram view)
►
Error counter is reset to 0. At the same time, the
counter for error-free measurements is also reset to 0. All counters can be shown via
<Settings><Results>.
3.7
Switching off
Move the on/off switch to the OFF position to switch off the flowscreen analysis system.