background image

 

Ver1.2  

                       

VPOD OPERATOR'S  MANUAL

 

 

Figure 2.2 rear view 

2.2 Install the batteries 

The oximeter can be powered by 2 AA-Size alkaline batteries (which will typically provide 50 hours of continuous operation), or by the optional rechargeable NiMH battery pack. 

When battery power is lower than 2.4V, the sign 

 

    will flicker in its display area. Replace the battery (or rechargeable NiMH battery pack) as soon as possible. The installation steps are shown as 

figure 2.3. 

 
Be sure to insert the batteries in the correct polarity, 

as indicated by polarity marking (+ and -) inside the battery compartment.   

 

Figure 2.3 

 

 

2.3 Connect the sensor 

Connect the oximeter sensor to the top of the oximeter as shown in figure 2.4. Ensure that the sensor is firmly plugged in. 

 

Figure2.4 

2.4 Monitoring 

Clip the sensor to the patient’s finger, and ensure that the patient’s nail surface is facing upward, as shown in figure 2.5. 

 

 

     

Fig2.5      Placement of the finger     

2.4.1 Power on the oximeter 
Press the function key (left key) to power it on. Several seconds later, the measurement value will appear. 

Notice: 

  To maintain the highest degree of accuracy, it is recommended that the finger and the oximeter sensor/probe is kept as still as possible.     

2.4.2 Brightness adjustment   

When you press the function key (left key) for more than one second, the brightness level headed with “Br” will be shown on the top right of the screen. You can adjust the brightness by degrees by 
pressing the setting key (right key). There are 10 levels of brightness. The default is level four. 

 

Figure 2.6

                                                                                                       

 

2.4.3 Mode switch 

After turning on the oximeter, each time the function key (left key) is pressed, the oximeter will switch to another display mode, shown as Figure 3.1.

 

 
2.5 Factors that may affect the measurement 

During operation, the accuracy of oximetry readings can be affected by the following factors: 

2.5.1  Instrument  performance  depends  on  the  pulsatile  character  of  the  artery.  The  measurement  would  not  be  considered  reliable  and  accurate  if  the  following  conditions  are  present  during 
measurement.     

  Shock or cardiac arrest 

  Temperature of the digit 

  After the administration of a cardiovascular drug 

  Anemia 

  Evidence of ventilation-perfusion mismatch 

2.5.2 Instrument performance depends on the wavelength absorption for oxyhemoglobin and deoxyhemoglobin. If there are substances absorbing the same wavelength, this would induce false or 
low SpO

2

 values. The following may affect these values:     

  carboxyhemoglobin 

  methemoglobin 

  methylene blue 

  Indigo carmine 

Содержание VPOD-H

Страница 1: ...ationwide Medicare approved Independent Diagnostic Testing Facility The VirtuOx Total Solution consists of our flagship Overnight Oximetry Testing platform VirtuOxTM now at NO CHARGE combined with our...

Страница 2: ...es depending on hemoglobin concentration The amount of light absorption at each light frequency is dependent on the degree of oxygenation of hemoglobin within the tissues The microprocessor can select...

Страница 3: ...It can be eliminated by covering each site with opaque material Optical cross talk may adversely affect the accuracy of the SpO2 readings Obstructions or dirt on the sensor s red light or detector ma...

Страница 4: ...the oximeter sensor probe is kept as still as possible 2 4 2 Brightness adjustment When you press the function key left key for more than one second the brightness level headed with Br will be shown...

Страница 5: ...tient motion conditions 3 2 Mode Introduction Power On Measurement and operation can be done normally There are three display modes Measure mode Information display mode and Trouble display mode 1 Dis...

Страница 6: ...ox software and a special data cable As for detailed setup and operation please refer to the Virtuox website www virtuox net Note Virtuox can be obtained at no charge with the purchase of this device...

Страница 7: ...with the operator s manual e The oximeter is installed or connected with such external device without our company permission as printer computer netline and lead to oximeter failure Our company will c...

Страница 8: ...rm Alarm Probe off Finger out Low battery Modes Visual Information Display Type OLED double color Parameters SpO2 PR Pleth waveform Pleth bar Mode 3 display modes Record Patient ID 10 patients Data re...

Страница 9: ...broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If...

Страница 10: ...Ver1 2 VPOD OPERATOR S MANUAL 9 objects and people...

Отзывы: