background image

 

Ver1.2  

                       

VPOD OPERATOR'S  MANUAL

 

 

Under the “Information Display mode”, press the function key (left key) for more than one second (extended press), the oximeter will enter into “Setting mode” and you will find a parameter item 
which include its title and value on the top right corner of the display. 

Under the setting mode, continually press “Function key (left key)”, the current parameter item will be changed sequentially in the following order: 

Br(brightness)-->ID(patientID)-->Y(year)-->M(month)-->D(date)-->H(hour)-->m(minute)-->S(second)--Br(brightness)-->... 

Notice: 

When setting, please don’t insert finger. 

b) Save and exit from “Setting mode” 

Under the “Setting mode”, press “function key (left key)” to select the desired parameter item.   

Then press “Setting key (right key)” to adjust the value. 

Each time when you press the “Setting key (right key)”, current parameter value will be added by 1 unit sequentially.   

Double press the “Setting key (right key)” and the current parameter setting will be added by 10 units sequentially. All use circular logic. 

To finish one setup, 

press

 

both left and right keys

 

together

 

to confirm

. Then the modification under the “Setting mode” will be saved, at the same time, the system will exit from “Setting 

mode”. 

Notice: 

 

Each time only can set one parameter item, you must redo the step 

  to 

  to finish every setup. 

 

The parameters adjustment only can be done in the “Setting mode” under “Information Display mode”.   

c) Cancel and exit from “Setting mode” 

Under the “Setting mode”, double-press the Function key (left key), the modification under “Setting mode” will be cancelled and simultaneously, the system will exit from “Setting mode”. 

If there is no operation under “Setting mode” for more than 10 seconds, the system will exit from “Setting mode” automatically.     

3.4 Data replay and transmission 

The oximeter can record SpO2 and PR value for more than 24 hours, and can analyze records one by one. You can transmit the history data to a PC using “Virtuox” software and a special data cable. 
As for detailed setup and operation, please refer to the “Virtuox” website.      www.virtuox.net 

Note: Virtuox can be obtained at no charge with the purchase of this device.      Please contact Virtuox customer service to set up an account.    877-337-7111 

CHAPTER 4 Maintenance and Repair   

4.1 Maintenance   

It is very important for user to perform daily maintenance of oximeter and parts in order to maintain its function and appearance. Disinfection procedures may be performed with the use of the below 
mentioned cleaner/disinfectants. Failure to perform these procedures may result in invalidating the warranty. Local disinfection protocols will apply. 
Please take out battery before cleaning the oximeter. 
 
The external surface of the oximeter can be cleaned by wiping with a damp cloth. Do not submerge the oximeter in any solution at any time. To do so will void the warranty 
Use the following permitted solutions: 

 

Ammonia (diluted) 

 

Glutaraldehyde   

 

10% Bleach solution 

 

Mild soapy water (diluted). 

 

Do not use the following cleaners: 

 Any kind of scrubbing or scouring solution 

 Acetone 

Alcohol-based cleaners 

       
Battery maintenance 

 

Remove the batteries if you will not be using the oximeter for an extended time. 

 

Charge the batteries (NiMH or Li) fully prior to storage. 

 

Please charge over 14 hours at first time, or will reduce the battery life. 

4.2 Troubleshooting 

a) Error Definitions 

Err 1: program memory damaged. 

Err 2: data memory damaged. 

Err 3: sensor

 

Red Emission Diode damaged. 

      Err 4: sensor Infrared-ray Emission Diode damaged.   

Err 5: sensor Infrared-ray Receipt Diode damaged. 

Err 6: exterior crystal oscillator damaged. 

Err 7: sensor emission diode or receipt diode damaged. 

Err 9: real time clock damaged. 

Err 10: EEPROM chip damaged. 

 

b) Possible problem and corresponding resolution 
 

Problems 

Possible reason 

Solution 

SpO

2

 or PR 

cannot be 

displayed 

normally 

1. Finger is not plugged 

correctly 

2. Patient’s Oxyhemoglobin 

value is too low to be 

measured 

1. Retry by plugging the finger 

2. Attempt several time to obtain a 

reading, If are sure that no 

problem exists, obtain further 

clinical examination 

Содержание VPOD-H

Страница 1: ...ationwide Medicare approved Independent Diagnostic Testing Facility The VirtuOx Total Solution consists of our flagship Overnight Oximetry Testing platform VirtuOxTM now at NO CHARGE combined with our...

Страница 2: ...es depending on hemoglobin concentration The amount of light absorption at each light frequency is dependent on the degree of oxygenation of hemoglobin within the tissues The microprocessor can select...

Страница 3: ...It can be eliminated by covering each site with opaque material Optical cross talk may adversely affect the accuracy of the SpO2 readings Obstructions or dirt on the sensor s red light or detector ma...

Страница 4: ...the oximeter sensor probe is kept as still as possible 2 4 2 Brightness adjustment When you press the function key left key for more than one second the brightness level headed with Br will be shown...

Страница 5: ...tient motion conditions 3 2 Mode Introduction Power On Measurement and operation can be done normally There are three display modes Measure mode Information display mode and Trouble display mode 1 Dis...

Страница 6: ...ox software and a special data cable As for detailed setup and operation please refer to the Virtuox website www virtuox net Note Virtuox can be obtained at no charge with the purchase of this device...

Страница 7: ...with the operator s manual e The oximeter is installed or connected with such external device without our company permission as printer computer netline and lead to oximeter failure Our company will c...

Страница 8: ...rm Alarm Probe off Finger out Low battery Modes Visual Information Display Type OLED double color Parameters SpO2 PR Pleth waveform Pleth bar Mode 3 display modes Record Patient ID 10 patients Data re...

Страница 9: ...broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If...

Страница 10: ...Ver1 2 VPOD OPERATOR S MANUAL 9 objects and people...

Отзывы: