TECHNICAL FEATURES 
2-1. Accuracy: ± (a% × reading data + digits), environment temperature at (23

°

C±5), relative humidity 

less than 75%,  
2-2. DC Voltage DCV 

Range Accuracy 

Resolution 

400mV ±(1.5%+6) 

0.1mV 

4V 1mV 
40V 10mV 
400V 

±(0.5%+6) 

100mV 

1000V ±(1.0%+4) 

1V 

Input impedance 400mV Range 10M

, other range is 40M

2-3. AC Voltage ACV 

Range Accuracy 

Resolution 

400mV ±(1.5%+6) 

0.1mV 

4V 1mV 
40V 10mV 
400V 

±(0.8%+6) 

100mV 

750V ±(1.0%+6) 

1V 

Input impedance 400mV Range 10M

, other range is 40M

. Frequency response: mV and 750V 

range: 40 to 100Hz, other range: 40 to 400 Hz. Display: Sine wave RMS (Average value response). 
2-4. DC Current DCA 

Range Accuracy 

Resolution 

400uA 0.1

µ

4000uA 1

µ

40mA 10

µ

400mA 

±(1.0%+5) 

100

µ

20A ±(1.2%+10)  10mA 

MAX measurement voltage drop: full range mA:1.2VA:100mV Max input current: 20A (within 15 
seconds) 
2-5. AC Current ACA 

Range Accuracy 

Resolution 

400uA 0.1

µ

4000uA 1

µ

40mA 10

µ

400mA 

±(1.5%+5) 

100

µ

20A ±(2.0%+15d)  10mA 

MAX measurement voltage drop: full range mA: 1.2V A:100mV Max input current: 20A (within 15 
seconds) Frequency response 20A range: 40 100Hz,other range:40 to 400Hz 
2-6. Resistance 

 

Range Accuracy 

Resolution 

400

 ±(0.8%+5d) 

0.1

 

4k

 

1

 

40k

 10

 

400k

 100

 

4M

 

±(0.8%+4d) 

1k

 

40M

 ±(1.2%+5d) 

10k

 

Open circuit voltage: 400mV 
NOTE: With a reading of less than 400 

, place the test leads together to measure the wire resistance, 

then subtract it from the actual measurement. 

Содержание VC88E

Страница 1: ...User s Manual for VC88E Digital Multimeter...

Страница 2: ...connection and the insulation of the test leads and probes to avoid electric shock 2 Never apply voltage higher than 1000VDC or 750VAC when measuring 3 Be careful when measuring a voltage higher than...

Страница 3: ...nse 2 4 DC Current DCA Range Accuracy Resolution 400uA 0 1 A 4000uA 1 A 40mA 10 A 400mA 1 0 5 100 A 20A 1 2 10 10mA MAX measurement voltage drop full range mA 1 2VA 100mV Max input current 20A within...

Страница 4: ...nce test Range Value displayed Testing condition Forward voltage drop of diode Forward DCA is approx 0 5mA the backward voltage is approx 1 5V Buzzer makes a long sound while resistance is less than 3...

Страница 5: ...uring function and range 4 Temperature positive and V Hz terminal 5 Voltage resistance frequency capacitor and temperature terminal 6 COM terminal 7 Less than 400mA current testing terminal and temper...

Страница 6: ...proper range based on the displayed value 2 If the LCD displays OL it means the current is over range You need to move the selector knob to a higher range 3 Maximum input current is 400mA or 20A base...

Страница 7: ...sert the black lest lead into the COM terminal and the red one into the V Hz terminal 2 Turn the selector knob to the Hz range 3 Place the test leads to the power source or the load which is to be tes...

Страница 8: ...the HOLD button again the HOLD function is cancelled 4 13 Auto power off 1 When the meter has not been used for about 15 minutes the built in buzzer will sound 5 times and the meter goes into sleeping...

Страница 9: ...ignal appears Replace battery Large Error Value Replace battery These specifications are subject to change without notice We hereby will not be responsible for any accident and damage caused by improp...

Отзывы: